Replacing the halogen lamp
Lamp Replacement and Calibration Corona extreme
4 Replacing the halogen lamp
4.1 General work
Warning
HAZARD FROM ELECTRICAL ENERGY
Make sure that the power supply is disconnected. The entire measuring
system must be disconnected from the power supply.
Make sure that the measuring system cannot be unintentionally reconnected
to the power supply by third parties.
Caution
GENERAL HAZARD
The activities below may only be performed by Service employees of
Carl Zeiss Spectroscopy GmbH or by appropriately trained personnel.
Caution
HAZARD FROM THERMAL ENERGY
Allow the halogen lamp to cool down for approx. 10 minutes before
removing it.
Caution
SENSITIVE OPTICAL COMPONENTS
Be careful not to touch the reflector and the glass bulb of the new halogen
lamp without cotton gloves.
Only carry out work on optical components in a very clean working
environment. Use gloves which are free of all traces of lint, powder or
grease.
Caution
SENSITIVE ELECTRONIC COMPONENTS
Perform all the activities below at an ESD workstation.