34
Lamp replacement and calibration - Corona extreme | Service Manual | EN_40_070_0011I
Replacing the halogen lamp
Lamp Replacement and Calibration Corona extreme
3
Disconnect the lamp cable from the adapter cable
Unplug the lamp cable plug
1
from the adapter cable
2
.
Fig. 36
1
2
Disconnecting the lamp cable from the adapter cable
4
Remove the halogen lamp
Unscrew the three Torx screws
1
from the lamp cap. Use a TX 10 Torx
screwdriver for this.
Carefully remove the old halogen lamp
2
upwards out of the reflector holder
3
.
Fig. 37
1
3
2
Removing the halogen lamp