12
Lamp replacement and calibration - Corona extreme | Service Manual | EN_40_070_0011I
Performing firmware update
Lamp Replacement and Calibration Corona extreme
3
Decide on further action
If the displayed
firmware version is < 151
, we recommend updating to
version 151.
⇒
“
Rev01_V0151.fwx”
In this case a batch file must be run.
⇒
“Install_Rev2_to_Rev3.bat” (see step 5 in this chapter)
If the displayed
firmware version is < 210
, we recommend updating to the
latest version.
⇒
“
V03.01_0210.fwx”
4
Perform firmware update
Note
IMPORTANT NOTE REGARDING FIRMWARE UPDATE
Do not switch off the device or interrupt the network connection tor the test
computer during the update under any circumstances.
Click on
Update device
under “Device info”.
Fig. 9 Performing firmware update
Select the correct firmware.
Hardware revision 1:
Hardware revision 2:
Hardware revision 3: