EN_40_070_0011I | Service Manual | Lamp replacement and calibration - Corona extreme
Replacing the halogen lamp
Lamp Replacement and Calibration Corona extreme
5
Install the lamp holder
Place the entire lamp holder – consisting of the upper lamp holder ring
2
and
lower lamp holder ring
3
together with the new halogen lamp
1
– carefully
on the optical shaft of the device. Ensure that the holes match OR are alligned
with the threaded holes.
Fasten the lamp holder by screwing the 2 Torx screws
4
back into the
threaded holes and tightening them.
Fig. 24
4
2
1
3
Inserting the lamp holder
6
Install new lamp cable
Insert the plug
1b
of the new lamp cable (from Maintenance Package 1) into
connector / jack
4
of the lamp board.
Insert the new lamp cable into the two cable clips
3
.
Close the two cable clips again.
Connect the plug
1a
with the new halogen lamp
2
.
Fig. 25
1a
2
4
1b
3
Installing new lamp cable
CAUTION
Make sure that the new halogen
lamp is precisely aligned.
Align the contacts of the lamp
exactly as shown in the figure to
the right (red line).