EN_40_070_0011I | Service Manual | Lamp replacement and calibration - Corona extreme
Performing sensor calibration
Lamp Replacement and Calibration Corona extreme
7
Select certificates
Enter your name in the “User” input field.
Use the
buttons to select your reference standard certificates.
Fig. 48 Selecting certificates
Note:
Make sure that the correct profile attachment is placed on the sensor.
Click
to proceed to the next step.
8
Referencing – Calibration of white reference standard
Place the white reference standard on the centering aid. Ensure correct
alignment (interlocking of groove and pin).
Click
OK
to start the calibration.
Fig. 49 Calibrating white reference standard
Remove the white reference standard when the measurement is complete.
“
Maximum certificate
”
= White reference standard
= SRS-99
“
Gray reference standard
”
= Gray reference standard
= SRS-60