To set the densitometer for use with Chromira Balance, first press the Menu button
on the upper right of the keypad. Push the Select Probe (left arrow) button five
times to scroll through the choices to the System Options menu. Press the Zero Cal
key (Yes) to enter the menu. Choose Yes when asked if you want to change Output
options, and enter the following settings, pressing the Read Enter button on the
lower left to switch between screens.
Output Type:
SERIAL:
Yes
PRINTER:
Yes/No
End-of-line characters:
CR:
Yes/No
LF:
Yes/No
Baud rate:
Don't care.
Data/Stop bits:
DATA:
7 / 8
STOP:
1 / 2
Serial print delay:
8
Send DEN/RGB:
DEN:
Yes/No
RGB:
Yes
Digits per channel:
3 / 4
Character between channels:
CR / LF / BLANK
Choose defaults on all other screens.
In the Chromira Balance application, press the settings button and adjust the baud
rate, data and stop bits to the selections made in the menu. Set the COM to the
port on which the densitometer is currently connected, and the Parity to
None
.