Frequency
The Digital ports may also be used for normal frequency (non-low level) measurements and
support frequencies up to 16KHz. Frequency readings typically take a few seconds to accurately
measure the incoming signal.
Event Counter
The Event Counter mode is normally used with rain gauge-type applications. However, the
counter can also be used with any sensor that acts as a switch closure or produces a 0 to 5 volt
pulse. The Slope (located under Math & Data Options) will probably need to be adjusted based
on the application (0.01 is typical for rain gauges, indicating 0.01 inches of rain). A total count
value is maintained until either the System 5000
is reset or the Value itself is manually reset
TM
through this screen.
5-8 Inputs
System 5000
TM