Wrong or Erratic Data:
1-
Check the sensor Properties screen for possible errors.
2-
Verify slopes, offsets, or custom functions.
3-
Check battery voltage and all connections.
SDI-12 Communications
No SDI-12 Response:
1-
Verify power to the System 5000
.
TM
2-
Check all SDI-12 connections, including DATA and GND.
3-
Check the pin out of the SDI-12 sensor for correct wiring.
4- Make sure all sensors have a unique address.
GOES Transmissions
GOES not transmitting:
1-
Check power to the GOES Radio, do so under a load (e.g. a transmission).
2-
See if the GOES time is correct, if it is stopped, it indicates power to the radio was lost,
stopping the clock. The radio will not transmit if the GOES clock is not running.
3-
Check all antenna wires.
4-
Check antenna alignment.
5-
Use a power meter to measure the power output of the radio and the reflected power.
System 5000
Maintenance and Troubleshooting 14-3
TM
Summary of Contents for WaterLog Series
Page 1: ...Model System 5000TM Owner s Manual Revision 1 4 2...
Page 2: ......
Page 8: ......
Page 24: ...2 10 Hardware Overview System 5000TM...
Page 58: ...5 14 Inputs System 5000TM...
Page 128: ...10 14 System Setup General Setup System 5000TM...
Page 144: ...12 6 System Setup Inputs Outputs System 5000TM...
Page 168: ...14 4 Maintenance and Troubleshooting System 5000TM...
Page 170: ...A 2 System 5000 Specifications System 5000TM...
Page 172: ...B 2 Pressure Sensor Option Module Specifications System 5000TM...
Page 174: ...C 2 Analog Digital Option Module Specifications System 5000TM...