4 to 20 Milliamp (4-20mA) Output
The 4-20mA output connection points are at the far right of the lower terminal strip.
Several instrumentation applications use sensors that provide an industry standard 4-20 milliamp
output signal. The System 5000
data logger / DCP can output a 4-20mA signal based on any of
TM
its inputs. For example, a user may want to connect a temperature probe to the System 5000
TM
and convert the temperature value into a 4-20mA output. A temperature probe on Analog
Channel 1 that produces a 0 to 5 volt output representing 0 to 100 degrees Celsius could easily be
setup to produce a 4-20 mA output that represents the 0 to 100 degrees. The System 5000
does
TM
not actually output a 4-20mA signal, but rather controls the current in an isolated externally
powered loop. The following figure shows a basic connection diagram.
2-6 Hardware Overview
System 5000
TM
Summary of Contents for WaterLog Series
Page 1: ...Model System 5000TM Owner s Manual Revision 1 4 2...
Page 2: ......
Page 8: ......
Page 24: ...2 10 Hardware Overview System 5000TM...
Page 58: ...5 14 Inputs System 5000TM...
Page 128: ...10 14 System Setup General Setup System 5000TM...
Page 144: ...12 6 System Setup Inputs Outputs System 5000TM...
Page 168: ...14 4 Maintenance and Troubleshooting System 5000TM...
Page 170: ...A 2 System 5000 Specifications System 5000TM...
Page 172: ...B 2 Pressure Sensor Option Module Specifications System 5000TM...
Page 174: ...C 2 Analog Digital Option Module Specifications System 5000TM...