4.10. Optional Sensors (External): CTD
The SeaBird MicroCat CTD with RS232 serial interface provides high quality conductivity, tem-
perature, salinity, and pressure (optional) data. When integrated with the ADV, MicroCat data is
collected at the beginning of each sampling interval. Data are integrated in a CTD data structure
within the ADV data set. See §3.9.2 for the commands used to control the CTD.
4.11. Optional Sensors (External): Analog Sensors
Depending on system configuration, the ADVField can have up to three available analog input
channels (0-4 VDC) for integrating a variety of sensors with analog output voltages. Common
applications include OBS or transmissometer data for suspended sediment studies. Data are sam-
pled once per second during the averaging interval; mean and standard deviation are also re-
corded. For information on available sensors, contact SonTek.
4.12. Optional Sensors (External): LISST-100 Sensors
The LISST-100 sensor is a Laser Interferometry and Transmissometry particle distribution meas-
urement device designed by Sequoia Scientific. This section describes the operational procedures
for using a LISST sensor with an ADVField/Hydra system. See §3.9.5 for the commands used to
control the LISST-to-ADVField interface.
4.12.1.
Download LISST Control Program
The
HYDRA.TTB
program will need to be loaded into the LISST-100. This must be done before
connecting the LISST to the ADVField/Hydra system. The loading process is simply a transmis-
sion of the ASCII characters to the LISST-100 with occasional pauses to allow the LIST-100
time to process the lines. The
NEW
command clears the current program from memory before the
new program is loaded. The
LIST
command can be used to print the currently loaded program to
the screen. The
TTOOLS
program provided by Sequoia Scientific is best suited for this download.
•
Connect LISST sensor to a serial port on your computer.
•
Connect external power source to the LISST (if applicable).
•
Start
TTOOLS
program (provided by Sequoia Scientific).
•
Choose the
Tattletale|Load Program
(Alt+L) option and choose the
HYDRA.TTB
file.
4.12.2.
Connect LISST to ADVField/Hydra
Attach the external sensor harness to the ADVField/Hydra electronics housing (8-pin male un-
derwater connector). Connect the LISST-100 sensor to the serial cable on the harness using the
LISST-CTD adaptor cable. The external cable harness is manufactured to accommodate the CTD
sensor. Because the CTD and the LISST are sharing the same serial port on the ADVField and
have a different pin-out, and adaptor cable is required to connect the LISST to the harness. Con-
nect the external power source to the LISST (if applicable).
Connect the ADVField to your computer and a power source using the ADVField power and
communications cable. On your computer, establish communications with the ADVField using
SONTERM
. In
SONTERM
, enable the LISST sensor using the LisstSensorInstalled command §3.9.5.
Verify that both the Paros and MicroCat CTD serial sensors are disabled. Note: The ParosFreq is
a frequency sensor and it can be operated simultaneously with the LISST.
Verify that you can communicate with the LISST via ADVField/Hydra COM port 1. To do this,
issue the
ECHO 1
command at the command prompt. In echo mode, you can send commands and
receive responses from the LISST sensor. To exit the echo mode, send a
BREAK
.
Summary of Contents for SonTek ADVField
Page 2: ......
Page 8: ...SonTek YSI iv ADVField Hydra Operation Manual September 1 2001 NOTES ...
Page 12: ......
Page 32: ...SonTek YSI 20 ADVField Operation Manual September 1 2001 ...
Page 102: ...SonTek YSI 90 ADVField Operation Manual September 1 2001 ...
Page 104: ...SonTek YSI 92 ADVField Operation Manual September 1 2001 ...
Page 118: ...SonTek YSI 106 ADVField Operation Manual September 1 2001 ...
Page 122: ...SonTek YSI 110 ADVField Operation Manual September 1 2001 ...
Page 126: ...SonTek YSI 2 ADVField Operation Manual September 1 2001 ...
Page 132: ...SonTek YSI vi ADVField Software Manual September 1 2001 NOTES ...
Page 136: ...SonTek YSI ADVField Software Manual September 1 2001 4 ...
Page 152: ...SonTek YSI ADVField Software Manual September 1 2001 20 ...
Page 160: ...SonTek YSI ADVField Software Manual September 1 2001 28 ...
Page 162: ...SonTek YSI ADVField Software Manual September 1 2001 30 ...
Page 164: ...SonTek YSI ADVField Software Manual September 1 2001 32 ...
Page 168: ...SonTek YSI ADVField Software Manual September 1 2001 36 ...
Page 174: ...SonTek YSI ADVField Software Manual September 1 2001 42 ...
Page 182: ...SonTek YSI ADVField Software Manual September 1 2001 50 ...
Page 184: ...SonTek YSI ADVField Software Manual September 1 2001 52 Figure 15 ViewHydra Screen Shot ...
Page 196: ...SonTek YSI ADVField Software Manual September 1 2001 64 ...
Page 198: ...SonTek YSI ADVField Software Manual September 1 2001 66 ...
Page 202: ......
Page 206: ......
Page 220: ...SonTek YSI Acoustic Doppler Velocimeter Principles of Operation September 1 2001 14 ...