SonTek/YSI
Using Frequency Pressure Sensors with SonTek/YSI Systems (December 2001)
8
5.5.
Windows Software Support
SonTek Windows-based software offers an alternative to the DOS extraction programs for han-
dling the external pressure sensor calibration data.
5.5.1. Downloading Calibration File Using SonUtils
The
Terminal
module within
SonUtils
provides an easy way to load the frequency pressure sen-
sor calibration file into the system processor. Simply establish communication with the system in
Terminal
, click the
Load Druck Calibration
icon, choose the appropriate file, and the load process
will be performed.
5.5.2. Specifying Calibration File When Using ViewHydra
When opening a data file (.adr) with
ViewHydra
, you must supply a sensor calibration file (.drk)
to convert pressure data to dbars. If no pressure calibration is provided, all pressure data will be
displayed in raw frequency counts.
ViewHydra
lets you save the calibration data in an associated
data file workspace, so the calibration data is loaded automatically the next time the data file is
opened.
After the data file is loaded, you can select the sensor offset from the processing menu to convert
the absolute pressure reading.
1. Run
DRUCK
/
PAROSFREQ
in
Terminal
and make note of the pressure reading with the system
in the air (say 10.025 dbar for this example).
2. In the
Processing|Sensor Calibration Settings…
menu, set
Sensor Type
to
Ext Pressure
, and
then enter a value of
–10.025
(for this example) in the
Offset
field to convert pressure data
into strain gage data (relative to the atmosphere).
3. Now use the
Processing|Recompute Burst Statistics File…
option to propagate the changes
into the burst-averaged data.
4. When exiting the program, use the
Save (Overwrite) Workspace
option to store the calibra-
tion and offset data. The program will now remember this information for this data set.
5.5.3. Using ViewADP with PCADP Data and Waves
Data collected by a standard ADP stores mean pressure data in dbars; no additional conversion is
needed. However, PC-ADP data and ADP wave series data (P-SERIES & PUV-Series) require
pressure calibration.
When opening a data file with
ViewADP
, click the
Load ExtPres CalFile
button to specify the
calibration file.
Notes:
1. If pressure calibration is not loaded, all pressure data will be displayed as raw frequencies.
2. If pressure calibration data is not loaded, wave data processing is disabled.
3.
ViewADP
will save the calibration data in the associated data file’s workspace. This allows
the calibration data to load automatically the next time the data file is opened.
Summary of Contents for SonTek ADVField
Page 2: ......
Page 8: ...SonTek YSI iv ADVField Hydra Operation Manual September 1 2001 NOTES ...
Page 12: ......
Page 32: ...SonTek YSI 20 ADVField Operation Manual September 1 2001 ...
Page 102: ...SonTek YSI 90 ADVField Operation Manual September 1 2001 ...
Page 104: ...SonTek YSI 92 ADVField Operation Manual September 1 2001 ...
Page 118: ...SonTek YSI 106 ADVField Operation Manual September 1 2001 ...
Page 122: ...SonTek YSI 110 ADVField Operation Manual September 1 2001 ...
Page 126: ...SonTek YSI 2 ADVField Operation Manual September 1 2001 ...
Page 132: ...SonTek YSI vi ADVField Software Manual September 1 2001 NOTES ...
Page 136: ...SonTek YSI ADVField Software Manual September 1 2001 4 ...
Page 152: ...SonTek YSI ADVField Software Manual September 1 2001 20 ...
Page 160: ...SonTek YSI ADVField Software Manual September 1 2001 28 ...
Page 162: ...SonTek YSI ADVField Software Manual September 1 2001 30 ...
Page 164: ...SonTek YSI ADVField Software Manual September 1 2001 32 ...
Page 168: ...SonTek YSI ADVField Software Manual September 1 2001 36 ...
Page 174: ...SonTek YSI ADVField Software Manual September 1 2001 42 ...
Page 182: ...SonTek YSI ADVField Software Manual September 1 2001 50 ...
Page 184: ...SonTek YSI ADVField Software Manual September 1 2001 52 Figure 15 ViewHydra Screen Shot ...
Page 196: ...SonTek YSI ADVField Software Manual September 1 2001 64 ...
Page 198: ...SonTek YSI ADVField Software Manual September 1 2001 66 ...
Page 202: ......
Page 206: ......
Page 220: ...SonTek YSI Acoustic Doppler Velocimeter Principles of Operation September 1 2001 14 ...