ADVField Software Manual (September 1, 2001)
14
Graphical Display
The main portion of the screen displays two plots. The upper plot represents the mean value for
signal-to-noise ratio (SNR) from the three receivers (in dB) and the mean value for correlation
for the three receivers (percent). The lower plot is the time-series of the three velocity compo-
nents. Both graphs are plotted left to right against time. A vertical bar identifies the current (most
recent) point. If the full span of the plot is reached (right end), plotting resumes on the left side.
If the ADV has the optional pressure sensor installed, the lower graph will include a fourth plot
for pressure data. This pressure data is plotted in units of counts with the axis set to full scale
(bottom of the plot corresponds to zero pressure, top corresponds to maximum pressure).
Active Keys
Several active keys (or key combinations) may be used during data collection. Note: Two of the
keys (Ctrl+S and Ctrl+Q) affect data recording while the program is in the data acquisition mode.
F1 - Display Probe
(Multiple choice)
Not used by ADVField systems.
F3 - Change Time Span
(Multiple choice)
This key lets you change the time span of the real-time trace from several preset alternatives.
F4 - Change Trace Vel
(Multiple choice)
This key lets you change which velocity components are plotted. You can display all three com-
ponents simultaneously or any single component.
F5 - Change Vel Scale
(Multiple choice)
This key lets you change the full velocity scale of the trace from several preset alternatives.
F6
-
Filter
(Multiple choice)
The data shown in the tabular display are filtered to allow you to estimate the mean data values
over a time scale that is determined by the filter setting. <F6> allows you to change the filter time
constant from several preset alternatives.
The filter used for the tabular display is a first-order recursive filter of the form:
y(i+1)=y(i)*a+x(i+1)*(1-a); where y(1) = x(1)
and where x(i) are the individual data samples from the ADV (for velocity, signal-to-noise ratio,
and correlation). The values y(i) are the mean values for each of these variables shown in the
tabular display. The standard deviation of velocity data is calculated as:
y2(i+1)=y2(i)*a+x(i+1)*x(i+1)*(1-a); where y2(1) = x(1)*x(1);
sqrt(y2(i+1)
-
y(i+1)*y(i+1))
If
dt
is the sampling interval (1/sampling frequency), and
Nfilt
is the total number of samples
within the filter width (Tfilt) (i.e., Nfilt=Tfilt/dt), then
a
=
(Nfilt-1)/Nfilt.
For example, for a 10-second filter when sampling at 25 Hz, a = 249/250. If a<0, a=0.
Summary of Contents for SonTek ADVField
Page 2: ......
Page 8: ...SonTek YSI iv ADVField Hydra Operation Manual September 1 2001 NOTES ...
Page 12: ......
Page 32: ...SonTek YSI 20 ADVField Operation Manual September 1 2001 ...
Page 102: ...SonTek YSI 90 ADVField Operation Manual September 1 2001 ...
Page 104: ...SonTek YSI 92 ADVField Operation Manual September 1 2001 ...
Page 118: ...SonTek YSI 106 ADVField Operation Manual September 1 2001 ...
Page 122: ...SonTek YSI 110 ADVField Operation Manual September 1 2001 ...
Page 126: ...SonTek YSI 2 ADVField Operation Manual September 1 2001 ...
Page 132: ...SonTek YSI vi ADVField Software Manual September 1 2001 NOTES ...
Page 136: ...SonTek YSI ADVField Software Manual September 1 2001 4 ...
Page 152: ...SonTek YSI ADVField Software Manual September 1 2001 20 ...
Page 160: ...SonTek YSI ADVField Software Manual September 1 2001 28 ...
Page 162: ...SonTek YSI ADVField Software Manual September 1 2001 30 ...
Page 164: ...SonTek YSI ADVField Software Manual September 1 2001 32 ...
Page 168: ...SonTek YSI ADVField Software Manual September 1 2001 36 ...
Page 174: ...SonTek YSI ADVField Software Manual September 1 2001 42 ...
Page 182: ...SonTek YSI ADVField Software Manual September 1 2001 50 ...
Page 184: ...SonTek YSI ADVField Software Manual September 1 2001 52 Figure 15 ViewHydra Screen Shot ...
Page 196: ...SonTek YSI ADVField Software Manual September 1 2001 64 ...
Page 198: ...SonTek YSI ADVField Software Manual September 1 2001 66 ...
Page 202: ......
Page 206: ......
Page 220: ...SonTek YSI Acoustic Doppler Velocimeter Principles of Operation September 1 2001 14 ...