SonTek/YSI
66
ADVField Operation Manual (September 1, 2001)
you will need to disconnect the cable from the underwater connector to the receiver board. This
cable is highly sensitive to noise and should not be modified.
To close the conditioning module, first inspect and clean all o-rings and o-ring surfaces. Replace
the o-rings if necessary. Next, connect the cable from the underwater connector to the receiver
board and make certain all cables from the optional sensors are securely connected (see §4.7.2 for
a description of these cables). The three screws that secure the end cap to the conditioning mod-
ule housing are slightly asymmetric and can only be installed in one orientation; this ensures the
correct alignment of the compass/tilt sensor with respect to the probe. Install the end cap and se-
cure with the three #6-32 screws.
Note: The cables and electronics in the conditioning module are very sensitive to noise and
should not be modified. If you have any questions, contact SonTek. When opening the
conditioning module, be sure to install a fresh bag of desiccant and/or purge the housing
with nitrogen before closing. This is particularly important in humid climates to prevent
condensation.
Summary of Contents for SonTek ADVField
Page 2: ......
Page 8: ...SonTek YSI iv ADVField Hydra Operation Manual September 1 2001 NOTES ...
Page 12: ......
Page 32: ...SonTek YSI 20 ADVField Operation Manual September 1 2001 ...
Page 102: ...SonTek YSI 90 ADVField Operation Manual September 1 2001 ...
Page 104: ...SonTek YSI 92 ADVField Operation Manual September 1 2001 ...
Page 118: ...SonTek YSI 106 ADVField Operation Manual September 1 2001 ...
Page 122: ...SonTek YSI 110 ADVField Operation Manual September 1 2001 ...
Page 126: ...SonTek YSI 2 ADVField Operation Manual September 1 2001 ...
Page 132: ...SonTek YSI vi ADVField Software Manual September 1 2001 NOTES ...
Page 136: ...SonTek YSI ADVField Software Manual September 1 2001 4 ...
Page 152: ...SonTek YSI ADVField Software Manual September 1 2001 20 ...
Page 160: ...SonTek YSI ADVField Software Manual September 1 2001 28 ...
Page 162: ...SonTek YSI ADVField Software Manual September 1 2001 30 ...
Page 164: ...SonTek YSI ADVField Software Manual September 1 2001 32 ...
Page 168: ...SonTek YSI ADVField Software Manual September 1 2001 36 ...
Page 174: ...SonTek YSI ADVField Software Manual September 1 2001 42 ...
Page 182: ...SonTek YSI ADVField Software Manual September 1 2001 50 ...
Page 184: ...SonTek YSI ADVField Software Manual September 1 2001 52 Figure 15 ViewHydra Screen Shot ...
Page 196: ...SonTek YSI ADVField Software Manual September 1 2001 64 ...
Page 198: ...SonTek YSI ADVField Software Manual September 1 2001 66 ...
Page 202: ......
Page 206: ......
Page 220: ...SonTek YSI Acoustic Doppler Velocimeter Principles of Operation September 1 2001 14 ...