ADVField/Hydra Operation Manual (September 1, 2001)
61
5. Connect the red, 2-pin input power connector to the Analog board. This cable goes to the
power and communication connector on the end plate.
6. Connect the red, 2-pin battery backup connector to the Analog board (systems with internal
recorder only). This cable goes to the backup battery mounted on the end plate.
7. Connect the black, 16-pin probe connector to the Analog board. This cable goes to the probe
connector on the end plate.
8. Connect the black, 10-pin temperature/pressure connector to the Analog board. This cable
goes to the probe connector on the end plate. This connector is
not
keyed; orient it such that
the two wires are connected to the pins farthest from the input power connector.
9. Connect the black, 16-pin auxiliary input/output connector to the DSP board. This cable goes
to the power and communication connector on the end plate.
10. Connect the black, 10-pin serial communication connector to the CPU board. This cable goes
to the power and communication connector on the end plate.
To close the canister:
1. Clean and inspect the o-ring and o-ring surfaces on the end plate and the canister. Replace the
o-ring if necessary.
2. Slide the end plate assembly into the canister by aligning the horizontal plate with the ex-
truded slots on the inside of the canister. Be careful that no cables or connectors catch on the
inside of the canister. The assembly should slide smoothly into the canister until the end plate
rests against the canister.
3. Secure the end plate using the four 1/4”-20 screws. Tighten securely (at least until the lock
washers have been completely flattened) without over tightening.
Summary of Contents for SonTek ADVField
Page 2: ......
Page 8: ...SonTek YSI iv ADVField Hydra Operation Manual September 1 2001 NOTES ...
Page 12: ......
Page 32: ...SonTek YSI 20 ADVField Operation Manual September 1 2001 ...
Page 102: ...SonTek YSI 90 ADVField Operation Manual September 1 2001 ...
Page 104: ...SonTek YSI 92 ADVField Operation Manual September 1 2001 ...
Page 118: ...SonTek YSI 106 ADVField Operation Manual September 1 2001 ...
Page 122: ...SonTek YSI 110 ADVField Operation Manual September 1 2001 ...
Page 126: ...SonTek YSI 2 ADVField Operation Manual September 1 2001 ...
Page 132: ...SonTek YSI vi ADVField Software Manual September 1 2001 NOTES ...
Page 136: ...SonTek YSI ADVField Software Manual September 1 2001 4 ...
Page 152: ...SonTek YSI ADVField Software Manual September 1 2001 20 ...
Page 160: ...SonTek YSI ADVField Software Manual September 1 2001 28 ...
Page 162: ...SonTek YSI ADVField Software Manual September 1 2001 30 ...
Page 164: ...SonTek YSI ADVField Software Manual September 1 2001 32 ...
Page 168: ...SonTek YSI ADVField Software Manual September 1 2001 36 ...
Page 174: ...SonTek YSI ADVField Software Manual September 1 2001 42 ...
Page 182: ...SonTek YSI ADVField Software Manual September 1 2001 50 ...
Page 184: ...SonTek YSI ADVField Software Manual September 1 2001 52 Figure 15 ViewHydra Screen Shot ...
Page 196: ...SonTek YSI ADVField Software Manual September 1 2001 64 ...
Page 198: ...SonTek YSI ADVField Software Manual September 1 2001 66 ...
Page 202: ......
Page 206: ......
Page 220: ...SonTek YSI Acoustic Doppler Velocimeter Principles of Operation September 1 2001 14 ...