Section 6. Compass Calibration (COMPCAL.EXE)
A program (
COMPCAL.EXE
) has been provided that lets you calibrate the optional compass/tilt
sensor. Calibration minimizes the effects of ambient magnetic fields on compass data. We rec-
ommend you perform a compass calibration before each ADV deployment.
COMPCAL
gives a graphical display of instrument orientation that can be used to ensure the com-
pass and tilt sensors are functioning correctly. To run the calibration program, make sure the in-
strument is powered up and connected to a computer serial port. Here is the command syntax.
COMPCAL [com port] [baud rate]
The
[com port]
parameter is the communication port being used on your computer (i.e.; 1 for
COM1, 2 for COM2);
[baud rate]
is the ADP communication rate (9600 is the default).
Once started,
COMPCAL
shows a graphical display of compass orientation. At this point, the com-
pass is not in calibration mode; it is in a monitoring mode where instrument orientation is output
in real-time. The graphical display provides three perspectives. To the right is a compass that in-
dicates the direction the X-axis is pointing. To the left are tabular and graphical displays of pitch
and roll. In the center is a 3D box with an X on the side corresponding to north. When this side is
facing into the screen, the instrument X-axis is facing north.
To begin the calibration, press <F1>.
COMPCAL
will put the compass in calibration mode, the dis-
play will resume, and instructions are provided for instrument rotation at the bottom of the
screen. The calibration procedure involves rotating the ADV slowly through two complete circles
while varying the pitch and roll as much as possible. Each turn should take about one minute; the
exact starting and ending orientations are not important.
To end the calibration, either press <F1> again (to remain in the graphical display) or press
<Esc> to exit with the calibration results. Pressing <Esc> at any time will exit the program and
display the results of the most recent calibration. See the
ADVField Operation Manual
for more
information about compass calibration and for information on how to interpret the calibration
score.
Summary of Contents for SonTek ADVField
Page 2: ......
Page 8: ...SonTek YSI iv ADVField Hydra Operation Manual September 1 2001 NOTES ...
Page 12: ......
Page 32: ...SonTek YSI 20 ADVField Operation Manual September 1 2001 ...
Page 102: ...SonTek YSI 90 ADVField Operation Manual September 1 2001 ...
Page 104: ...SonTek YSI 92 ADVField Operation Manual September 1 2001 ...
Page 118: ...SonTek YSI 106 ADVField Operation Manual September 1 2001 ...
Page 122: ...SonTek YSI 110 ADVField Operation Manual September 1 2001 ...
Page 126: ...SonTek YSI 2 ADVField Operation Manual September 1 2001 ...
Page 132: ...SonTek YSI vi ADVField Software Manual September 1 2001 NOTES ...
Page 136: ...SonTek YSI ADVField Software Manual September 1 2001 4 ...
Page 152: ...SonTek YSI ADVField Software Manual September 1 2001 20 ...
Page 160: ...SonTek YSI ADVField Software Manual September 1 2001 28 ...
Page 162: ...SonTek YSI ADVField Software Manual September 1 2001 30 ...
Page 164: ...SonTek YSI ADVField Software Manual September 1 2001 32 ...
Page 168: ...SonTek YSI ADVField Software Manual September 1 2001 36 ...
Page 174: ...SonTek YSI ADVField Software Manual September 1 2001 42 ...
Page 182: ...SonTek YSI ADVField Software Manual September 1 2001 50 ...
Page 184: ...SonTek YSI ADVField Software Manual September 1 2001 52 Figure 15 ViewHydra Screen Shot ...
Page 196: ...SonTek YSI ADVField Software Manual September 1 2001 64 ...
Page 198: ...SonTek YSI ADVField Software Manual September 1 2001 66 ...
Page 202: ......
Page 206: ......
Page 220: ...SonTek YSI Acoustic Doppler Velocimeter Principles of Operation September 1 2001 14 ...