SonTek/YSI
Using Frequency Pressure Sensors with SonTek/YSI Systems (December 2001)
6
5. Additional Software
This section describes additional software required to use external frequency pressure sensors
(
DRUCK
/
PAROSFREQ
) with SonTek instruments.
5.1.
Downloading and Extracting Druck Calibration: LDDRUCK/GTDRUCK
The
LDDRUCK.EXE
program
loads
the
DRUCK
/
PAROSFREQ
calibration file into the instrument’s
processor. The command syntax is:
LDDRUCK [DruckCalFile] [-COM port] [-Baud Rate] [-Address]
where
•
DruckCalFile
is the name of the calibration file (no extension).
•
-COM port
is the serial port to which the system is connected (1=COM1; 2= COM2).
•
-Baud Rate
is the communication baud rate setting of the system. If no value is given, the
program assumes the default setting of 9600 baud. Acceptable baud rate settings are 1200,
2400, 4800, 9600, 19200, 38400, 57600, and 115200.
•
-Address
is the address for a RS-485 system.
Example:
LDDRUCK P84567 –p1 –b19200 –a2
will download the file
P84567.DRK
onto the
system at address
2
connected to COM port
1
at
19200
baud.
The
GTDRUCK.EXE
program
gets
(downloads) the
DRUCK
/
PAROSFREQ
calibration file from the in-
strument’s processor to the user’s PC. The command syntax is:
GTDRUCK [MyCalFile] [-COM port] [-Baud Rate] [-Address]
where
•
MyCalFile
is the name of the calibration file (no extension).
•
-COM port
is the serial port to which the system is connected (1=COM1; 2= COM2).
•
-Baud Rate
is the communication baud rate setting of the system. If no value is given, the
program assumes the default setting of 9600 baud. Acceptable baud rate settings are 1200,
2400, 4800, 9600, 19200, 38400, 57600, and 115200.
•
-Address
is the address for a RS-485 system.
Example:
GTDRUCK B199 –p2 –b19200
will retrieve the calibration data from the system
connected to COM port
2
at
19200
baud and store the data into file
B199.DRK
.
Summary of Contents for SonTek ADVField
Page 2: ......
Page 8: ...SonTek YSI iv ADVField Hydra Operation Manual September 1 2001 NOTES ...
Page 12: ......
Page 32: ...SonTek YSI 20 ADVField Operation Manual September 1 2001 ...
Page 102: ...SonTek YSI 90 ADVField Operation Manual September 1 2001 ...
Page 104: ...SonTek YSI 92 ADVField Operation Manual September 1 2001 ...
Page 118: ...SonTek YSI 106 ADVField Operation Manual September 1 2001 ...
Page 122: ...SonTek YSI 110 ADVField Operation Manual September 1 2001 ...
Page 126: ...SonTek YSI 2 ADVField Operation Manual September 1 2001 ...
Page 132: ...SonTek YSI vi ADVField Software Manual September 1 2001 NOTES ...
Page 136: ...SonTek YSI ADVField Software Manual September 1 2001 4 ...
Page 152: ...SonTek YSI ADVField Software Manual September 1 2001 20 ...
Page 160: ...SonTek YSI ADVField Software Manual September 1 2001 28 ...
Page 162: ...SonTek YSI ADVField Software Manual September 1 2001 30 ...
Page 164: ...SonTek YSI ADVField Software Manual September 1 2001 32 ...
Page 168: ...SonTek YSI ADVField Software Manual September 1 2001 36 ...
Page 174: ...SonTek YSI ADVField Software Manual September 1 2001 42 ...
Page 182: ...SonTek YSI ADVField Software Manual September 1 2001 50 ...
Page 184: ...SonTek YSI ADVField Software Manual September 1 2001 52 Figure 15 ViewHydra Screen Shot ...
Page 196: ...SonTek YSI ADVField Software Manual September 1 2001 64 ...
Page 198: ...SonTek YSI ADVField Software Manual September 1 2001 66 ...
Page 202: ......
Page 206: ......
Page 220: ...SonTek YSI Acoustic Doppler Velocimeter Principles of Operation September 1 2001 14 ...