SonTek/YSI
ADVField Software Manual (September 1, 2001)
29
Section 5. Analog Output Calibration (ADFCALAO.EXE)
All ADVField systems generate analog output voltages. The analog outputs include three lines
for the three velocity components, and one line proportional to the mean signal amplitude. Sam-
pling these voltage lines with an external analog data acquisition system provides an alternative
means of recording ADV data. When possible, we recommend using the digital velocity data as
this provides improved resolution (avoiding a digital-to-analog-to-digital conversion) and addi-
tional diagnostic data (three signal amplitude and three correlation values). See the
ADVField
Operation Manual
for detailed information about the analog output voltages.
The ADV software includes a program (
ADFCALAO.EXE
) for calibrating the analog output volt-
ages. Calibration of the analog outputs is required to compensate for minor changes in the output
voltage levels caused by cables and temperature fluctuations. To run the calibration program,
type the following command at the DOS prompt.
ADFCALAO [com port]
The
[com port]
parameter is the communication port to which the instrument is connected
(1 for COM1, 2 for COM2).
ADFCALAO
accesses the ADV and forces all analog outputs to their low, mid, or high-level volt-
ages. The low voltage corresponds to maximum negative velocity; mid to zero velocity; and high
to maximum positive velocity. The operator sets the level by simply pressing the L, M, or H key.
After pressing one of these keys, the program will set the analog outputs of the ADV to the cor-
responding level, so the actual output voltage can be measured by a data acquisition system or a
precision voltmeter. See
ADVField Operation Manual
for details on applying the output voltage
calibration.
To exit the program, press <Esc>.
Summary of Contents for SonTek ADVField
Page 2: ......
Page 8: ...SonTek YSI iv ADVField Hydra Operation Manual September 1 2001 NOTES ...
Page 12: ......
Page 32: ...SonTek YSI 20 ADVField Operation Manual September 1 2001 ...
Page 102: ...SonTek YSI 90 ADVField Operation Manual September 1 2001 ...
Page 104: ...SonTek YSI 92 ADVField Operation Manual September 1 2001 ...
Page 118: ...SonTek YSI 106 ADVField Operation Manual September 1 2001 ...
Page 122: ...SonTek YSI 110 ADVField Operation Manual September 1 2001 ...
Page 126: ...SonTek YSI 2 ADVField Operation Manual September 1 2001 ...
Page 132: ...SonTek YSI vi ADVField Software Manual September 1 2001 NOTES ...
Page 136: ...SonTek YSI ADVField Software Manual September 1 2001 4 ...
Page 152: ...SonTek YSI ADVField Software Manual September 1 2001 20 ...
Page 160: ...SonTek YSI ADVField Software Manual September 1 2001 28 ...
Page 162: ...SonTek YSI ADVField Software Manual September 1 2001 30 ...
Page 164: ...SonTek YSI ADVField Software Manual September 1 2001 32 ...
Page 168: ...SonTek YSI ADVField Software Manual September 1 2001 36 ...
Page 174: ...SonTek YSI ADVField Software Manual September 1 2001 42 ...
Page 182: ...SonTek YSI ADVField Software Manual September 1 2001 50 ...
Page 184: ...SonTek YSI ADVField Software Manual September 1 2001 52 Figure 15 ViewHydra Screen Shot ...
Page 196: ...SonTek YSI ADVField Software Manual September 1 2001 64 ...
Page 198: ...SonTek YSI ADVField Software Manual September 1 2001 66 ...
Page 202: ......
Page 206: ......
Page 220: ...SonTek YSI Acoustic Doppler Velocimeter Principles of Operation September 1 2001 14 ...