SonTek/YSI
100
ADVField Operation Manual (September 1, 2001)
RecordCompass Set to YES
If
RecordCompass
=
YES
, an additional six bytes of data are recorded with each sample (assumes
the optional compass/tilt sensor is installed).
Datum
Bytes Description
Heading
2
Heading angle in units of 0.1 degree
Pitch
2
Pitch angle in units of 0.1 degree
Roll
2
Roll angle in units of 0.1 degree
==
6
RecordSensor Set to YES
If
RecordSensor
=
YES
, an additional four bytes of data are recorded with each sample (assumes
the optional temperature and/or pressure sensor is installed).
Datum
Bytes Description
Temperature
2
Temperature in units of 0.01°C
Pressure*
2
Pressure in units of counts (0-65535)
==
*
Note: If
ExtSensor
=
3_OBS
, raw Ad Chan 3 data is displayed here.
4
RecordExtSensor Set to YES
If
RecordExtSensor
=
YES
, an additional four bytes of data are recorded with each sample (as-
sumes at least one optional external sensor is installed). The data corresponds to raw counts from
a 12-bit analog-to-digital converter.
Datum
Bytes Description
Ad Chan 1
2
Data from sensor channel #1 in units of counts (0-65535)
Ad Chan 2
2
Data from sensor channel #2 in units of counts (0-65535)
==
4
RecordExtPress Set to YES
If
RecordExtPress
=
YES
, an additional three bytes of data are recorded with each sample (as-
sumes an optional external pressure sensor is installed).
Datum
Bytes Description
ExtPressure
3
Data from external pressure sensor in milli-dbars (for
==
Paros) or milli-Hz (for Druck and ParosFreq)
3
RecordCTD Set to YES
If
RecordCTD
=
YES
, an additional four bytes of data are recorded in a single record at the end of
the burst (i.e., not at the end of each sample). This data contains the mean temperature, con-
ductivity, pressure, and salinity measured by an external MicroCat CTD throughout the burst.
Datum
Bytes Description
Temp
1
Mean temperature from CTD sensor in 0.0001°C
Cond
1
Mean conductivity from CTD sensor in 0.00001 Siemen
Press
1
Mean pressure from CTD sensor in 0.001 dbar
Sal
1
Mean salinity from CTD sensor in 0.0001 psu
==
4
Summary of Contents for SonTek ADVField
Page 2: ......
Page 8: ...SonTek YSI iv ADVField Hydra Operation Manual September 1 2001 NOTES ...
Page 12: ......
Page 32: ...SonTek YSI 20 ADVField Operation Manual September 1 2001 ...
Page 102: ...SonTek YSI 90 ADVField Operation Manual September 1 2001 ...
Page 104: ...SonTek YSI 92 ADVField Operation Manual September 1 2001 ...
Page 118: ...SonTek YSI 106 ADVField Operation Manual September 1 2001 ...
Page 122: ...SonTek YSI 110 ADVField Operation Manual September 1 2001 ...
Page 126: ...SonTek YSI 2 ADVField Operation Manual September 1 2001 ...
Page 132: ...SonTek YSI vi ADVField Software Manual September 1 2001 NOTES ...
Page 136: ...SonTek YSI ADVField Software Manual September 1 2001 4 ...
Page 152: ...SonTek YSI ADVField Software Manual September 1 2001 20 ...
Page 160: ...SonTek YSI ADVField Software Manual September 1 2001 28 ...
Page 162: ...SonTek YSI ADVField Software Manual September 1 2001 30 ...
Page 164: ...SonTek YSI ADVField Software Manual September 1 2001 32 ...
Page 168: ...SonTek YSI ADVField Software Manual September 1 2001 36 ...
Page 174: ...SonTek YSI ADVField Software Manual September 1 2001 42 ...
Page 182: ...SonTek YSI ADVField Software Manual September 1 2001 50 ...
Page 184: ...SonTek YSI ADVField Software Manual September 1 2001 52 Figure 15 ViewHydra Screen Shot ...
Page 196: ...SonTek YSI ADVField Software Manual September 1 2001 64 ...
Page 198: ...SonTek YSI ADVField Software Manual September 1 2001 66 ...
Page 202: ......
Page 206: ......
Page 220: ...SonTek YSI Acoustic Doppler Velocimeter Principles of Operation September 1 2001 14 ...