SonTek/YSI
ADVField/Hydra Operation Manual (September 1, 2001)
5
•
The optional pressure sensor is installed on the upper end cap of the conditioning module,
near the underwater cable connector. Note that because the pressure sensor is mounted some
distance from the sampling volume, this distance may need to be taken into account when
processing the combined pressure and velocity data.
•
The optional temperature sensor (a thermistor) is mounted inside the upper end cap of the con-
ditioning module, next to the underwater cable connector on probes that have a sampling dis-
tance of 5 cm. When using this type of temperature sensor, be certain that the conditioning
module is completely submerged. On probes that have a sampling distance of 10 cm, the tem-
perature sensor is potted inside the acoustic sensor head.
•
The high-frequency cable to the processing module carries analog signals from the probe to
the digital processing electronics. This cable is highly sensitive to noise and should not be
modified without instruction from SonTek.
•
The signal-conditioning module has a maximum diameter of 10.1 cm (4.0 in) and is 33.0 cm
(13.0 in) long. The outside of the housing is ribbed for easier handling and more secure
mounting. The overall dimensions of the probe vary with 2D and 3D probes, distance to the
sampling volume, stem length, and sensor orientation.
1.1.3. 5-MHz ADVOcean Probe
The 5-MHz ADVOcean probe configurations are divided into two groups: with or without op-
tional sensors (compass/tilt, pressure). The 5-MHz ADVOcean probe is used for field applica-
tions where a rugged probe that is resistant to physical damage is needed. Figure 5 shows the
standard ADVOcean probe. The ADVOcean probe is made of stainless steel. However, the con-
ditioning module can be made of stainless steel or plastic if weight is a concern.
5-MHz ADVOcean Probe Configuration without Optional Sensors
:
•
The ADVOcean probe consists of an acoustic sensor head (which includes a temperature sens-
ing thermistor), a signal conditioning module, and an underwater mateable connector to a
high-frequency cable. The probe serial number is engraved on the housing near the acoustic
transmitter. All ADVOcean probes having a stainless steel conditioning module include a sac-
rificial zinc anode attached to the signal-conditioning module for corrosion protection. Probes
with plastic conditioning modules require you to provide appropriate protection, if applicable.
•
The acoustic sensor head contains three acoustic receivers and one acoustic transmitter. The
sensor head is mounted directly to the end of the signal-conditioning module.
Figure 5 – Standard 5-MHz ADVOcean Probe
Acoustic receivers (3)
Acoustic transmitter
Acoustic sensor
head assembly
Sampling
volume
Signal
conditioning
module
Underwater
mateable
connector
High-frequency cable
(to Processor)
Summary of Contents for SonTek ADVField
Page 2: ......
Page 8: ...SonTek YSI iv ADVField Hydra Operation Manual September 1 2001 NOTES ...
Page 12: ......
Page 32: ...SonTek YSI 20 ADVField Operation Manual September 1 2001 ...
Page 102: ...SonTek YSI 90 ADVField Operation Manual September 1 2001 ...
Page 104: ...SonTek YSI 92 ADVField Operation Manual September 1 2001 ...
Page 118: ...SonTek YSI 106 ADVField Operation Manual September 1 2001 ...
Page 122: ...SonTek YSI 110 ADVField Operation Manual September 1 2001 ...
Page 126: ...SonTek YSI 2 ADVField Operation Manual September 1 2001 ...
Page 132: ...SonTek YSI vi ADVField Software Manual September 1 2001 NOTES ...
Page 136: ...SonTek YSI ADVField Software Manual September 1 2001 4 ...
Page 152: ...SonTek YSI ADVField Software Manual September 1 2001 20 ...
Page 160: ...SonTek YSI ADVField Software Manual September 1 2001 28 ...
Page 162: ...SonTek YSI ADVField Software Manual September 1 2001 30 ...
Page 164: ...SonTek YSI ADVField Software Manual September 1 2001 32 ...
Page 168: ...SonTek YSI ADVField Software Manual September 1 2001 36 ...
Page 174: ...SonTek YSI ADVField Software Manual September 1 2001 42 ...
Page 182: ...SonTek YSI ADVField Software Manual September 1 2001 50 ...
Page 184: ...SonTek YSI ADVField Software Manual September 1 2001 52 Figure 15 ViewHydra Screen Shot ...
Page 196: ...SonTek YSI ADVField Software Manual September 1 2001 64 ...
Page 198: ...SonTek YSI ADVField Software Manual September 1 2001 66 ...
Page 202: ......
Page 206: ......
Page 220: ...SonTek YSI Acoustic Doppler Velocimeter Principles of Operation September 1 2001 14 ...