SonTek/YSI
ADVField/Hydra Operation Manual (September 1, 2001)
71
standoffs; board orientation does not matter. Above the sensor board are four standoffs (#4-40 x
1/2”). Last is the compass, secured by four #4-40 x 3/8” bolts, with fiber washers on each side.
Align the compass as shown in Figure 21. For side-looking operation, it is not possible to align
the compass X-axis with the probe X-axis. Instead, the probe X-axis is aligned with the compass
negative Y-axis. Assuming the orientation switch is set for side-looking operation (§4.2), this
alignment is accounted for internally.
In side-looking operation, you must
ensure the probe’s X-axis (stamped
on the end cap and defined by re-
ceiver arm #1) is in the horizontal
plane (within the
±
50
°
limits of the
tilt sensor). When looking at the rear
end cap, the pressure sensor should
be in the position shown in Figure 22.
If the system does not have a pressure
sensor, make sure the X-axis points
to the right of the underwater connec-
tor. Figure 22 shows the orientation.
If the system is not oriented correctly,
the compass will be upside down and
will give invalid data.
Probe X
X
Z
Compass Axis
Compass/Tilt
Sensor
3/8" Hex Standoffs
Pressure / Temp
Sensor Board
X
Y
Compass Axis
Figure 21 - Compass/Tilt Sensor, Side-Looking Mounting
Receiver #1
Vertically Up
Pressure
Underwater
M
bl
Connector
Figure 22 - Probe Mounting, Side-Looking Compass
Summary of Contents for SonTek ADVField
Page 2: ......
Page 8: ...SonTek YSI iv ADVField Hydra Operation Manual September 1 2001 NOTES ...
Page 12: ......
Page 32: ...SonTek YSI 20 ADVField Operation Manual September 1 2001 ...
Page 102: ...SonTek YSI 90 ADVField Operation Manual September 1 2001 ...
Page 104: ...SonTek YSI 92 ADVField Operation Manual September 1 2001 ...
Page 118: ...SonTek YSI 106 ADVField Operation Manual September 1 2001 ...
Page 122: ...SonTek YSI 110 ADVField Operation Manual September 1 2001 ...
Page 126: ...SonTek YSI 2 ADVField Operation Manual September 1 2001 ...
Page 132: ...SonTek YSI vi ADVField Software Manual September 1 2001 NOTES ...
Page 136: ...SonTek YSI ADVField Software Manual September 1 2001 4 ...
Page 152: ...SonTek YSI ADVField Software Manual September 1 2001 20 ...
Page 160: ...SonTek YSI ADVField Software Manual September 1 2001 28 ...
Page 162: ...SonTek YSI ADVField Software Manual September 1 2001 30 ...
Page 164: ...SonTek YSI ADVField Software Manual September 1 2001 32 ...
Page 168: ...SonTek YSI ADVField Software Manual September 1 2001 36 ...
Page 174: ...SonTek YSI ADVField Software Manual September 1 2001 42 ...
Page 182: ...SonTek YSI ADVField Software Manual September 1 2001 50 ...
Page 184: ...SonTek YSI ADVField Software Manual September 1 2001 52 Figure 15 ViewHydra Screen Shot ...
Page 196: ...SonTek YSI ADVField Software Manual September 1 2001 64 ...
Page 198: ...SonTek YSI ADVField Software Manual September 1 2001 66 ...
Page 202: ......
Page 206: ......
Page 220: ...SonTek YSI Acoustic Doppler Velocimeter Principles of Operation September 1 2001 14 ...