NOTE CAREFULLY: Users who are accustomed to installing a Teflon membrane on a YSI dissolved
oxygen sensor should note that the methods recommended for the Teflon installation are not likely to work
well for the free chlorine membrane because this membrane cannot be stretched. Therefore, dissolved
oxygen membrane installation techniques which are outline in the manuals for various YSI instruments,
including standard 6-series sondes utilizing the 6562 Rapid Pulse DO sensor, are NOT RECOMMENDED
for the free chlorine membrane installation.
2.3.2 STEP 2 - INSTALLING THE PROBES
Remove the calibration cup from your sonde by hand as shown below to expose the bulkhead.
REMOVING THE PORT PLUGS
Using the long extended end of the probe installation tool supplied in the YSI 6570
Maintenance Kit, remove the port plugs. Save all the port plugs
for possible future use.
TRANSPORT CUP
BULKHEAD WITH
PROBE PORT PLUGS