650 MDS
Section 3
YSI, Incorporated Drinking Water Monitoring Systems Operations Manual
3-32
At the end of the 90 minute profile determination, the user either confirms the
Stop logging
command by
pressing
Enter
or simply presses the
Escape
key to return to the sonde
Discrete sample
menu, highlights the
Close file
selection, and confirms the file closing by pressing
Enter
.
The profile data taken at a 16 second sample interval is now stored in the sonde memory under the file name
PROFILE1. It can be viewed on the 650 display and/or uploaded to the 650 memory or directly to a PC by
accessing the sonde
File
menu.
3.5.2.4
3.5.2.4
3.5.2.4
3.5.2.4
SONDES WITHOUT MEMORY
If your sonde is a 600R, 600XL, or 6820 purchased prior to September 1999, it may contain a circuit board
that does not have memory. With sondes of this type, you will not be able to log data to the sonde – only to
the 650 as detailed in the following section. With these no-memory sondes attached to your 650, the
Run
display will show a blank Sonde logging window as shown below.