650 MDS
Section 3
YSI, Incorporated Drinking Water Monitoring Systems Operations Manual
3-23
3.4.2 SONDE MENU E
3.4.2 SONDE MENU E
3.4.2 SONDE MENU E
3.4.2 SONDE MENU EXAMPLES
XAMPLES
XAMPLES
XAMPLES
To become familiar with the
650 Sonde menu
function, YSI recommends that the user connect a sonde to
their 650 and proceed through the following two examples of common sonde interface operations.
Example 1: Setting up the Report Output – Activation of the Salinity Parameter
Highlight the
Sonde menu
selection in the 650 Main menu and press
Enter
. Then use the arrow keys
highlight the
Report
selection and press
Enter
again.
Use the arrow keys to highlight the
Salinity
selection and then press the
Enter
key to activate the
parameter as shown by a dot in the open circle. Note that the entire report list can be accessed by
continuing to scroll the highlighting block down with the arrow key. After activation of salinity in the
Report
setup of the sonde, press the
Escape
key twice in succession to return to the
Main sonde
menu.
Example 2: Calibrating the Dissolved Oxygen Sensor
NOTE: In the following example, it is assumed that the
Autosleep RS232
function of your sonde has been
disabled in the sonde
Advanced|Setup
menu. If
Autosleep
were active, then the display would indicate a
countdown of the DO warm-up time and calibrate automatically at the end of this time period. See Section
3.4.9 of this manual and Section 2 of your 6-series manual for more information on the
Autosleep
function.
Place the sonde in the calibration chamber with the dissolved oxygen probe in water-saturated air.
Highlight the
Sonde menu
selection in the 650 Main menu and press
Enter
. Then, use the arrow keys to
highlight the
Calibration
selection and press
Enter
.