IM 253401-01E
4-3
4
Setting Measurement Conditions
4.2
Turning the Filter ON/OFF
Relevant Keys
SCALING
AVG
FILTER
STORE
RECALL
HARMONICS
SAMPLE
V OVER
A OVER
MODE
RMS
V MEAN
DC
A
B
C
hour
hour
min
min
sec
V
VA
m
A
k
var
M W
TIME
V
PF
m
A
k
deg
M W
%
FUNCTION
AUTO
AUTO
MODE
1
Φ
3W
V
Hz
m
A
k
h
M W
TRIG
V RANGE A RANGE
HOLD
ENTER
INTEGRATOR
START
HARMONICS
MEMORY
INTEG SET
STOP
RESET
REMOTE
INTERFACE
OUTPUT
LOCAL
SETUP
h
SHIFT
WIRING
3
Φ
4W
3
Φ
3W
3V3A
ELEMENT
1
2
3
FUNCTION
ELEMENT
1
2
3
FUNCTION
ELEMENT
1
2
3
*Shows the operation panel of the WT130. For the differences
between WT110 and WT130, refer to section 2.2, page 2-2, 2-3
Displays
relevant
keys and
indicator
Operating Procedure
Selecting the filter
(Display C)
ENTER
3.
SETUP
1.
ENTER
(Display C)
4.
2.
5
.
End
• Perform operations following the thick line in the below menu.
• Press the ENTER key to confirm the selection or setting.
• When you want to leave the current menu during operation, press the key described
under step 1. The confirmed settings made until that point will be kept.
Explanation
Filter Function
The instrument will perform measurements after synchronizing to the cycle of the input signal.
Consequently, the frequency of the input signal can be measured properly. The filter, at a cut-
off frequency of 300Hz, will only be applied to the frequency measurement circuit and will
remove noise from distorted and inverted waves, etc.. This allows the frequency to be measured
correctly which improves the accuracy of each measurement value. The filter will not be applied
to the voltage and current circuit. The initial value is OFF.
Note
The filter setting cannot be changed while integration is being carried out.