Appendix 2.3 Commands
Communication Commands 2
2.3.10 RECall Group
The commands in the RECall group are used to make settings relating to, and inquires about recalling data. This allows you to
make the same settings and inquiries as can be set using the lower menus of [MEMORY]-”rECAL” or [MEMORY]-”PnLrC”.
:RECall
:
STAT e
<Space>
OFF
ON
<NRf>
?
INTer val
<Space>
<NRf>
,
<NRf>
,
<NRf>
?
PANel
<Space>
<NRf>
?
<Character string>
;
RECall?
Function
Queries all the settings relating to recalling data.
Syntax
RECall?
Example
RECALL?
→
:RECALL:STATE 0;INTERVAL 0,0,0
RECall:INTerval
Function
Sets the recalling interval/queries the current setting.
Syntax
RECall:INTerval {<NRf>,<NRf>,<NRf>|<String>}
RECall:INTerval?
{<NRf>,<NRf>,<NRf>}=0,0,0 to 99,59,59
{<String>}=HH:MM:SS
HH
hour
MM
minutes
SS
seconds
Example
RECALL:INTERVAL 0,0,0
RECALL:INTERVAL “00:00:00“
RECALL:INTERVAL?
→
:RECALL:INTERVAL 0,0,0
Description
Even when the interval has been set to 0,0,0, the
interval becomes 250ms in case of normal
measurement and 1s in case of harmonic analysis.
RECall:PANel
Function
Retrieves the set-up parameters file.
Syntax
RECall:PANel {<NRf>}
{<NRf>}=1 to 4 : file number
Example
RECALL:PANEL 1
RECall[:STATe]
Function
Turns recalling ON/OFF, queries the current setting.
Syntax
RECall[:STATe] {<Boolean>}
RECall:STATe?
Example
RECALL:STATE ON
RECALL:STATE?
→
:RECALL:STATE 1