IM 253401-01E
10-18
Explanation
Setting the Output (Printing) Mode
This setting is to select whether you are printing out on a plotter or a printer.
HPGL : For printing on an external, HPGL - compatible plotter.
ESCP : For printing on an external, ESC / P - compatible printer.
PCL
: For printing on an external, PCL5 (printer language of HP) - compatible printer. This
mode is available on WT110/WT130 with version 2.21 or later.
Setting the Output Contents in case of Normal Measurement
All measured / computed data will be output.
Setting the Output Items and the Element in case of Harmonic Analysis
• Setting the Output Item (Column A)
One of the following items should be set, which then will be printed out on an external plotter/
ptinter. The initial value is V.
V
: Prints the numerical values of the analysis value and relative harmonic content of the
voltage;
A
: Prints the numerical values of the analysis value and relative harmonic content of the
current;
P
: Prints the numerical values of the analysis value and relative harmonic content of the
active power;
dEG
: Prints the numerical values of the phase angle;
G-V
: Prints the numerical values
*1
and the graph of the analyzed voltage values;
G-A
: Prints the numerical values
*1
and the graph of the analyzed current values;
G-P
: Prints the numerical values
*1
and the graph of the analyzed active power values;
G-Vd : Prints the numerical values
*1
and the graph of the phase angle between each voltage of
the 2nd to 50th order and the fundamental (1st order);
G-Ad : Prints the numerical values
*1
and the graph of the phase angle between each current of
the 2nd to 50th order and the fundamental (1st order);
CG-V : Prints the numerical values
*1
and the graph of the relative harmonic content of voltage;
CG-A : Prints the numerical values
*1
and the graph of the relative harmonic content of current;
CG-P : Prints the numerical values
*1
and the graph of the relative harmonic content of active
power;
ALL : Prints the numerical values
*1
and the graph of the analysis values and relative
harmonic content of voltage and current (V and A are both printed).
*1 HPGL/PCL plotters print both numerical values and the graph, but ESCP printers only print
the graph.
• Setting the Element (Column B)
One of the following should be set. The output items corresponding to the set element will then
be printed out on an external plotter. The initial value is 1. In case of the WT110, this setting is
always 1.
1 : Select this when the output items of element 1 should be printed out;
2 : Select this when the output items of element 2 should be printed out; This setting is not
available on model 253502.
3 : Select this when the output items of element 3 should be printed out.
Executing Output
After having connected the external plotter / printer to this instrument, execute the output of data.
dATA : All data selected as output items will be output.
PnL
: All set-up parameters will be output.
10.9 Outputting to an External Plotter / Printer