IM 253401-01E
App2-16
Appendix 2.3 Commands
[CONFigure]:SCALing?
Function
Queries all settings relating to the scaling function.
Syntax
[CONFigure]:SCALing?
Example
[CONFIGURE]:SCALING?
→
:CONFIGURE:
SCALING:STATE 0;PT:ELEMENT1 1.000E+00;
ELEMENT2 1.000E+00;ELEMENT3 1.000E+00;
:CONFIGURE:SCALING:CT:
ELEMENT1 1.000E+00;ELEMENT2 1.000E+00;
ELEMENT3 1.000E+00;:CONFIGURE:SCALING:
SFACTOR:ELEMENT1 1.000E+00;
ELEMENT2 1.000E+00;ELEMENT3 1.000E+00
[CONFigure]:SCALing:{PT|CT|SFACtor}?
Function
Queries all scaling values related to
{voltage|current|power}
.
Syntax
[CONFigure]:SCALing:{PT|CT|SFACtor}?
Example
[CONFIGURE]:SCALING:PT?
→
:CONFIGURE:SCALING:
PT:ELEMENT1 1.000E+00;ELEMENT2 1.000E+00;
ELEMENT3 1.000E+00
[CONFigure]:SCALing:{PT|CT|SFACtor}[:ALL]
Function
Sets the scaling values for all elements of
{voltage|current|power}
at once.
Syntax
[CONFigure]:SCALing:{PT|CT|SFACtor}[:ALL]
{<NRf>}
{<NRf>}=0.001 to 1000
Example
[CONFIGURE]:SCALING:PT:ALL 1.000
Description
The setting values differ as follows.
Less than 1.000 : Three digits after the decimal point
are valid.
1.000 to 1000 : The first five digits are valid.
[CONFigure]:SCALing:{PT|CT|SFACtor}:
ELEMent<x>
Function Sets the scaling value for
{voltage|current|power}
of each element, queries the current setting.
Syntax
[CONFigure]:SCALing:{PT|CT|SFACtor}:
ELEMent<x> {<NRf>}
[CONFigure]:SCALing:{PT|CT|SFACtor}:
ELEMent<x>?
<x>= 1
(WT110 single-phase model)
1, 3
(WT130 three-phase, three-wire
model)
1 to 3 (WT130 three-phase, four-wire
model)
{<NRf>}=0.001 to 1000
Example
[CONFIGURE]:SCALING:PT:ELEMENT1 1.000
[CONFIGURE]:SCALING:PT:ELEMENT1?
→
:
CONFIGURE:SCALING:PT:ELEMENT1 1.000E+00
Description
Setting values differ as described at
[CONFigure]:CURRent:ESCaling[:ALL]
.
[CONFigure]:SCALing[:STATe]
Function
Sets scaling ON/OFF, queries the current setting.
Syntax
[CONFigure]:SCALing[:STATe] {<Boolean>}
[CONFigure]:SCALing:STATe?
Example
[CONFIGURE]:SCALING:STATE OFF
[CONFIGURE]:SCALING:STATE?
→
:CONFIGURE:
SCALING:STATE 0
[CONFigure]:VOLTage?
Function
Queries all settings relating to voltage range.
Syntax
[CONFigure]:VOLTage?
Example
[CONFIGURE]:VOLTAGE?
→
:CONFIGURE:VOLTAGE:
RANGE 600.0E+00;AUTO 1
[CONFigure]:VOLTage:AUTO
Function
Sets the voltage auto range ON/OFF, queries the
current setting.
Syntax
[CONFigure]:VOLTage:AUTO {<Boolean>}
[CONFigure]:VOLTage:AUTO?
Example
[CONFigure]:VOLTage:AUTO ON
[CONFIGURE]:VOLTAGE:AUTO?
→
:CONFIGURE:
VOLTAGE:AUTO 1
[CONFigure]:VOLTage:RANGe
Function
Sets the voltage range/queries the current setting.
Syntax
[CONFigure]:VOLTage:RANGe {<voltage>}
[CONFigure]:VOLTage:RANGe?
<voltage>=15V to 600V (15, 30, 60, 150, 300,
600V)
Example
[CONFIGURE]:VOLTAGE:RANGE 600V
[CONFIGURE]:VOLTAGE:RANGE?
→
:CONFIGURE:
VOLTAGE:RANGE 600.0E+00
[CONFigure]:WIRing
Function
Sets the wiring method/queries the current setting.
Syntax
[CONFigure]:WIRing {P1W2|P1W3|P3W3|P3W4|V3A3}
[CONFigure]:WIRing?
Example
[CONFIGURE]:WIRING P1W3
[CONFIGURE]:WIRING?
→
:CONFIGURE:WIRING P1W3
Description
The selections stand for the following.
P1W2 : Single-phase, two-wires ( only for WT110)
P1W3 : Single-phase, three-wires (only for WT130)
P3W3 : Three-phase, three-wires (only for WT130)
P3W4 : Three-phase, four-wires (only for WT130 3-
phase, 4-wire model)
V3A3 : Three-voltage, three-current (only for WT130
3-phase, 4-wire model)