IM 253401-01E
11-5
11
GP-IB Interface
Output Format when Self Selected
Up to 14 normal measured/computed data can be output simultaneously, and the user is allowed
to choose any output information type for those 14 data. Each output block is of the following
format.
ch.1
,
ch.2
ch.3
ch.4
Terminator
,
,
ch.5
,
ch.6
ch.7
ch.8
Terminator
,
,
ch.9
,
ch.10
ch.11
ch.12
Terminator
,
,
ch.13
,
ch.14
Terminator
END
Terminator
Line 1
Line 2
Line 3
Line 4
Line 5
Line 6
Data
number
Terminator
(The data number will only be output in case of recall)
Each output block usually consists of five lines (six in case of recall) including the block end
line “END”. However, if all output types on a line are set to “no output”, this line will be
omitted, reducing the number of output lines by one. For example, if all output items of ch.9 to
ch12 are set to “no output”, line 4 in the above example will be omitted.
Furthermore, if any channel on a line is set to “no output”, all data following this channel on the
line will be shifted forward. For example, if the ch.2 on line 1 is set to “no output”, data of ch.1
will be followed by data of ch.3.
Output Format in case of Normal Measurement
WT110 (253401)
V1 data
Terminator
A1 data
Terminator
W1 data
Terminator
Frequency
,
Display C
Terminator
END
Terminator
Line 1
Line 2
Line 3
Line 4
Line 5
Line 6
Data
number
Terminator
(The data number will only be output in case of recall)
WT130 (253502)
V1 data
,
V3 data
SV data
Terminator
,
A1 data
,
A3 data
SA data
Terminator
,
W1 data
,
W3 data
SW data
Terminator
,
Frequency
,
Display C
Terminator
END
Terminator
Line 1
Line 2
Line 3
Line 4
Line 5
Line 6
Data
number
Terminator
(The data number will only be output in case of recall)
Note
• When the frequency is set by either of the following methods, only one value is measured, and that value
will be output.
- by panel keys : by the FUNCTION key and ELEMENT key of display C (except WT110)
- by communication command : by the “DC” or “EC” command.
After setting the measurement object of frequency, even changing the display C to something different
than VHz or AHz will not result in changing the object of measurement of frequency. When selecting the
output items yourself and you set a frequency item which is not object of measurement, “999999.E+03”
will be output.
11.4 Output Format for Normal Measured/Computed Data, Harmonic Analysis Data, Set-up Parameters and Error Codes