IM 253401-01E
App2-44
Appendix 2.7 Communication-related Error Messages
Errors in communications execution (200 to 299)
Code
Message
Action
Reference Page
221
Setting conflict
Check the relevant setting.
App.2.3
222
Data out of range
Check the setting range.
App.2.3
223
Too much data
Check the data byte length.
App.2.3
224
Illegal parameter value
Check the setting range.
App.2.3
241
Hardware missing
Check availability of options.
—
260
Expression error
Equation is not allowed.
—
270
Macro error
Does not conform to the macro definition specified in IEEE488.2.
—
272
Macro execution error
Does not conform to the macro definition specified in IEEE488.2.
—
273
Illegal macro label
Does not conform to the macro definition specified in IEEE488.2.
—
275
Macro definition too long
Does not conform to the macro definition specified in IEEE488.2.
—
276
Macro recursion error
Does not conform to the macro function specified in IEEE488.2.
—
277
Macro redefinition not allowed
Does not conform to the macro definition specified in IEEE488.2.
—
278
Macro header not found
Does not conform to the macro definition specified in IEEE488.2.
—
Error in communication Query(400 to 499)
Code
Message
Action
Reference Page
410
Query INTERRUPTED
Check transmission/reception order.
App2-3
420
Query UNTERMINATED
Check transmission/reception order.
App2-3
430
Query DEADLOCKED
Limit the length of the program message including <PMT> to
App2-4
1024 bytes or less.
440
Query UNTERMINATED after
Do not enter any query after *IDN? and *OPT?.
—
indefinite response
Errors in Execution (800 to 899)
Codes
Message
Action
Reference Page
813 to 819
Invalid operation
Refer to 14.4
14-11
830 to 833
Internal memory access error
Refer to 14.4
14-11
841 to 847
Integrator execute error
Refer to 14.4
14-11
Error in System Operation (912)
Code
Message
Action
Reference Page
912
Fatal error in Communication driver Service is required.
—
Warnings (350, 390)
Code
Message
Action
Reference Page
350
Queue overflow
Read out the queue.
App2-39
390
Overrun error
Adjust the baud rate.
—
(only for RS-232C)
Note
The warning code 350 only appears in case of an overflow of the error queue. The error which occurs in case
of clearing the
STATus:ERRor?