4-12
IM 704420-01E
4.6
Setting Inhibit
«For a functional description, see section 1.4.»
Keys
KEY
LOCK
MEAS
SET
AVE
/T
0
2
4
4
8
6
12
8
16
10
%
deg
%
20
180
us
ms
ns
MEAS
x1
3TCD
xN
3TCD
xN
BI-PHASE
SPEED
DELAY
GATE TIME
FUNC
GATE
LOCAL
DATA
SCALE
PRESET
PHASE
10
20
TRIG
CLOCK
PHASE
AUTO MAN
GP-IB
REMOTE
EQLZ
LEVEL
PLL
SHIFT
ARM
INH
DtoC
DVD
0.1s
0.5s
MAN
UTILITY
OPTION
10
5
σ
σ
Procedure
Turning ON inhibit and selecting the polarity
Press the INH key to select
or
. The indicator of the selected item lights and
inhibit is turned ON.
Turning OFF inhibit
Press the INH key to turn OFF both the
and
indicators. Inhibit is turned OFF.
Explanation
You can inhibit measurements by applying an external signal (inhibit signal) to the
INHIBIT input connector. This is possible even while the gate is open or during
measurement after arming activation.
Selecting the polarity
•
: Inhibits measurements while a positive signal is being input.
•
: Inhibits measurements while a negative signal is being input.
Inhibit signal
Input the inhibit signal to the connector indicated as “INHIBIT” on the rear panel.
Item
Specifications
Input impedance
10 k
Ω
(typical value*)
Input coupling
DC
Input level
TTL level
Allowable input voltage range
–8 V to 13 V (DC+ACpeak)
Minimum input pulse width
30 ns
Setup time
0 ns (possible even when the inhibit signal and data signal
are simultaneous)
*
The typical value is a representative or standard value. It is not a warranted value.