8-4
IM 704420-01E
GP-IB Cable
The GP-IB connector used on this instrument is a 24-
pin connector that conforms to the IEEE St’d 488.1-
1978. Use a GP-IB cable that conforms to this
standard.
Connection Procedure
Connect the cable as shown below.
GP-IB connector
GP-IB cable
Precautions to Be Taken When
Connecting the Cable
• Firmly tighten the screws on the GP-IB cable
connector.
• Multiple devices can be connected to a single GP-IB
system. However, no more than 15 devices
(including the controller) can be connected to a
single system.
• When connecting multiple devices, each device
must have its own unique address.
• Use a cable of length 2 m or less for connecting the
devices.
• Make sure the total cable length does not exceed 20
m.
• When communicating, have at least two-thirds of
the devices turned ON.
• When connecting multiple devices, connect them in
a star or linear configuration (see the diagram
below). Loop and parallel configurations are not
allowed.
8.3
Connecting the GP-IB Cable