7-2
IM 704420-01E
Setting the determination level
12.
Press the SHIFT+<(UTILITY) key. The display shows the characters [init].
13.
Turn the rotary knob to select [dcJdG].
14.
Press the > key. The display shows the determination level.
15.
Use the rotary knob and arrow keys (< or >) to set the determination level.
For the procedure to set numeric values, see section 3.7.
Setting the average coefficient of the DC output filter
16.
Press the SHIFT+<(UTILITY) key. The display shows the characters [init].
17.
Turn the rotary knob to select [AVE].
18.
Press the > key. The display shows the average coefficient of the DC output
filter.
19.
Use the rotary knob and arrow keys (< or >) to set the average coefficient.
For the procedure to set numeric values, see section 3.7.
Setting the jitter ratio correction coefficient
α
20.
Press the SHIFT+<(UTILITY) key. The display shows the characters [init].
21.
Turn the rotary knob to select [ALPHA].
22.
Press the > key. The display shows the correction coefficient
α
.
23.
Use the rotary knob and arrow keys (< or >) to set the jitter ratio correction
coefficient
α
.
For the procedure to set numeric values, see section 3.7.
Setting the jitter ratio correction coefficient
β
24.
Press the SHIFT+<(UTILITY) key. The display shows the characters [init].
25.
Turn the rotary knob to select [bEtA].
26.
Press the > key. The display shows the correction coefficient
β
.
27.
Use the rotary knob and arrow keys (< or >) to set the jitter ratio correction
coefficient
β
.
For the procedure to set numeric values, see section 3.7.
If you press the MEAS key in the middle or at the end of the setting operation, the
information that is specified up to that point is applied to the DC output setting and the
measurement operation resumes.
7.1 Setting the DC Output