8-34
IM 704420-01E
8.7.17 Common Command Group
The commands in the common group are defined in the IEEE 488.2-1992 and are independent of the
instrument’s functions.
;
*CAL
?
?
<Space>
<NRf>
*ESE
?
?
?
?
?
<Space>
<NRf>
*CLS
*ESR
?
*IDN
*OPC
*OPC
*TRG
*RST
*SRE
*STB
*TST
*WAI
8.7 Commands
* C A L ?
* C A L ?
* C A L ?
* C A L ?
* C A L ?
Function
Performs calibration and queries the result.
Syntax
*CAL?
Example
*CAL?
Description If the calibration terminates normally, “
0
” is
returned. If abnormality is detected, a non-zero
value is returned. There is another calibration
execution command. For details, see page 9-
12.
* C L S
* C L S
* C L S
* C L S
* C L S
Function
Clears the standard event register, extended
event register, and error queue.
Syntax
*CLS
Example
*CLS
Description For details on the standard event register,
extended event register, and error queue, see
section 8.8, “Status Report.”
* E S E
* E S E
* E S E
* E S E
* E S E
Function
Sets the standard event enable register or
queries the current setting.
Syntax
*ESE <NRf>
*ESE?
<NRF>=0 to 255
Example
*ESE 253
*ESE?
→
253
Description For details on the standard event enable
register, see section 8.8, “Status Report.”
* E S R ?
* E S R ?
* E S R ?
* E S R ?
* E S R ?
Function
Queries the standard event register and clears
the register.
Syntax
*ESR?
Example
*ESR?
→
253
Description For details on the standard event register, see
section 8.8, “Status Report.”
* I D N ?
* I D N ?
* I D N ?
* I D N ?
* I D N ?
Function
Queries the instrument model.
Syntax
*IDN?
Example
*IDN?
→
YOKOGAWA,704420,0,F1.01
Description A reply is returned in the following form:
manufacturer, model, serial number (always 0),
firmware version.
* O P C
* O P C
* O P C
* O P C
* O P C
Function
Sets a “
1
” to the standard event register bit upon
the completion of the specified overlap
command. Because the instrument does not
support overlap commands, the command is
discarded.
Syntax
*OPC
* O P C ?
* O P C ?
* O P C ?
* O P C ?
* O P C ?
Function
Returns a “
1
” when the specified overlap
command is finished. A “
1
” is always returned,
because the instrument does not support
overlap commands.
Syntax
*OPC?