9-3
IM 704420-01E
Troubleshooting and Maintenance
9
Code
Message
Description and Corrective Action
Section
148
Character data not allowed
<CHARACTER DATA> cannot be used. Write in
8.7
a data form other than {...|...|...}.
150
String data error
There is no delimiter to the right of
8.6
<STRING DATA>.
Enclose <String data> in double quotation or
single quotation marks.
151
Invalid string data
The contents of <STRING DATA> are
8.7
inappropriate. <String data> is too long or
invalid character is present.
158
String data not allowed
<STRING DATA> cannot be used. Write in a
8.7
data form other than <String data> form.
161
Invalid block data
The data length of <BLOCK DATA> does not match. 8.6, 8.7
<Block data> cannot be used.
168
Block data not allowed
<BLOCK DATA> cannot be used.
8.6, 8.7
<Block data> cannot be used.
171
Invalid expression
There is an invalid character in the
8.7
<EXPRESSION DATA>.
Equations cannot be used.
178
Expression data not allowed
<EXPRESSION DATA> cannot be used.
8.7
Equations cannot be used.
181
Invalid outside macro definition
The placeholder is outside the macro.
—
Macro functions defined in IEEE488.2 are not
supported.
Error in communication execution
Code
Message
Description and Corrective Action
Section
221
Setting conflict
There is a conflict in the setup information.
8.7
Check the relevant setting values.
222
Data out of range
The value of <DATA> is outside the range.
8.7
Check the range.
223
Too much data
The length of <DATA> is too long. Check the 8.7
length of the data.
224
Illegal parameter value
The value of <DATA> is inappropriate. Check 8.7
the range.
241
Hardware missing
The hardware is not implemented. Check the
—
existence of options.
260
Expression error
<EXPRESSION DATA> is not correct.
—
Equations cannot be used.
270
Macro error
Macro nesting is too deep. Macro functions
—
defined in IEEE488.2 are not supported.
272
Macro execution error
Macros cannot be used. Macro functions
—
defined in IEEE488.2 are not supported.
273
Illegal macro label
The macro label is inappropriate. Macro
—
functions defined in IEEE488.2 are not
supported.
275
Macro definition too long
The macro is too long. Macro functions
—
defined in IEEE488.2 are not supported.
276
Macro recursion error
Macro was recursively called. Macro
—
functions defined in IEEE488.2 are not
supported.
277
Macro redefinition not allowed
Macros cannot be redefined. Macro functions —
defined in IEEE488.2 are not supported.
278
Macro header not found
The macro is not defined. Macro functions
—
defined in IEEE488.2 are not supported.
708
Cannot output data while
Cannot transmit data while measurement is
8.7.11,
measuring.
in progress. Stop the measurement using the 8.7.14
STOP or SSTART command.
9.2 Error Code Description and Corrective Actions