7-1
IM 704420-01E
Outputting Signals, Initializing Setup Information, and Setting Key Lock
7
7.1
Setting the DC Output
«For a functional description, see section 1.6.»
Connecting the Cable
Connect a BNC cable to the jitter DC output connector on the rear panel of the TA120E.
Keys
When setting the DC output mode, DC output
range of the jitter ratio, determination level,
and average coefficient, characters or values
indicating the settings are displayed.
Lights when the numeric values are shown on the display.
KEY
LOCK
MEAS
SET
AVE
/T
0
2
4
4
8
6
12
8
16
10
%
deg
%
20
180
us
ms
ns
MEAS
x1
3TCD
xN
3TCD
xN
BI-PHASE
SPEED
DELAY
GATE TIME
FUNC
GATE
LOCAL
DATA
SCALE
PRESET
PHASE
10
20
TRIG
CLOCK
PHASE
AUTO MAN
GP-IB
REMOTE
EQLZ
LEVEL
PLL
SHIFT
ARM
INH
DtoC
DVD
0.1s
0.5s
MAN
UTILITY
OPTION
10
5
σ
σ
Procedure
Selecting the DC output mode
1.
Press the SHIFT+<(UTILITY) key. The display shows the characters [init].
2.
Turn the rotary knob to select [dcout].
3.
Press the > key. Turn the rotary knob to select [Jitt] or [JudGE]. If you select
[Jitt] (jitter ratio output), a DC voltage corresponding to the jitter ratio is output
from the jitter DC output connector. If you select [JudGE] (determination
output), a DC voltage of 5 V is output when the jitter ratio is below the
determination level and 0 V when the jitter is above the determination level from
the jitter DC output connector.
Setting the jitter ratio output range
• Setting the upper limit of the jitter ratio output range
4.
Press the SHIFT+<(UTILITY) key. The display shows the characters [init].
5.
Turn the rotary knob to select [dc hi].
6.
Press the > key. The display shows the upper limit of the jitter ratio output
range.
7.
Use the rotary knob and arrow keys (< or >) to set the upper limit.
For the procedure to set numeric values, see section 3.7.
• Setting the lower limit of the jitter ratio output range
8.
Press the SHIFT+<(UTILITY) key. The display shows the characters [init].
9.
Turn the rotary knob to select [dc Lo].
10.
Press the > key. The display shows the lower limit of the jitter ratio output
range.
11.
Use the rotary knob and arrow keys (< or >) to set the lower limit.
For the procedure to set numeric values, see section 3.7.
Chapter 7 Outputting Signals, Initializing Setup Information, and Setting Key Lock