3-17
IM PX8000-02EN
3.14 Triggering on a Wave Window Trigger
(Enhanced)
This section explains the following settings for triggering on a wave window trigger.
• Source channel
Tolerance width, cycle frequency, and reference cycle
• Synchronization channel
• Trigger condition
►
Features Guide: “Wave Window Trigger (Enhanced)”
ENHANCED Wave Window Trigger Menu
Press
SIMPLE/ENHANCED
and then the
Setting
soft key to select Enhanced. The following menu
appears.
Set Setting to Enhanced.
Set the trigger hold off. ► section 3.3
Set Type to Wave Window.
Set the trigger conditions.
Setting the Trigger Conditions (Set Pattern)
Press the
Set Pattern
soft key to display the following menu.
Set the source channel (ON, OFF).
Set the tolerance width.
Set the cycle frequency.
Set the reference cycle.
Set the level for detecting the start
and end points, and set the detection
hysteresis.*
* Set when the synchronization channel is U1 to U4, I1 to I4, P1 to P4, or AUX3 to AUX8.
Set the synchronization channel
(Auto, U1-U4, I1-I4, P1-P4, AUX3-AUX8).