25-4
IM PX8000-02EN
Execution Errors (650 to 799)
Code
Message
Chapter or
Section
650
Data is invalid.
—
651
The option is not equipped, so it cannot execute.
—
652
Undo is not possible since data that existed immediately before initialization is not available.
—
653
Can not be executed while running.
Press START/STOP key to stop acquisition.
4.2
654
Cannot manipulate files while image printing is in progress.
Wait until image printing is complete.
—
656
Calibration failure.
Disconnect the input and execute again.
If it fails again, servicing is necessary.
24.1
660
Can not operate while data out.
Wait until output is completed.
—
663
Cannot start.
—
664
GO/NO-GO is available while trigger mode is
- ‘Single’ or ‘Normal’
- ‘Auto’ or ‘AutoLevel’ (Faster than 50ms/div)
3.1
666
Failed to measure statistics.
Waveform data may be missing.
If Cycle Statistics is specified, the instrument may be configured in a way that fails to detect the cycle.
14.2, 14.3
667
Executing file access.
Abort or wait until it is complete.
—
668
Image is being printed or saved.
Wait until the execution of the command is complete.
—
672
Cannot be executed when the time base setting is to be an External clock.
4.1
674
Average practice can’t be done because the record length of the history exceeds the record length that
it can be carried out.
20.1
675
Average practice can’t be done because the record length of the history exceeds the record length that
it can be carried out.
20.1
677
Cannot do while selftest is executing.
25.2
679
Cannot start at the current record length. Shorten the record length or meet the following condition.
- Set the trigger mode to Auto, decrease T/Div to less than 100 msec/div to enable roll mode.
- Set the trigger mode to Single or On Start.
2.8, 3.1
680
Averaging mode is not possible when the trigger mode is Single, SingleN, or On Start.
Change the trigger mode.
3.1
684
Cannot start when the time base set to external clock while the acquisition mode set to envelope.
4.1
686
Cannot be executed when the acquisition mode is set to average.
Change the mode.
4.1
693
Cannot be executed when GO/NO-GO mode is Zone.
17.1
695
Set acquisition mode to Normal when using a wave window trigger.
4.1
696
The wave window trigger cannot be used if the sampling rate is faster than 500 kS/s or slower than 10
kS/s.
2.8
702
All search conditions are OFF.
At least one condition should ON.
Chapter 19,
section 20.2
703
Display setting of search source is OFF.
Set it to ON.
8.1
704
Cannot execute Time search while T/div is faster than 100msec/div.
2.8
705
Cannot start Action mode while trigger mode is SingleN.
3.1
706
Cannot be executed when GO/NO-GO mode is ON.
Set the GO/NO-GO mode to OFF.
Chapter 17
707
Cannot execute search while searched No. reached Maximum(1000).
—
708
Cannot execute or set while AutoScroll processing.
Stop AutoScroll.
12.1
712
Cannot start while No GO/NO-GO condition.
Chapter 17
713
Cannot make wave zone from less than 2,000 points data, from more than 10,000,000 points data, or
from less than 10division data.
—
714
Cannot start Action mode while PrintImage target is “File”.
Change target to “printer”.
21.2
716
Set the Math and FFT Window to OFF to Start GO/NO-GO.
Chapter 15,
section 16.1
717
Cannot abort this process.
—
25.1 Messages and Corrective Actions