IM 12B07D02-01E
Troubleshooting 8-3
Table 8-1. Error Codes
Code Error description
Possible cause
Suggested remedy
E0
Buffer solution temperature
Buffer solution too hot or too cold
Adjust buffer temperature
outside the programmed range
Check cabling
E1
Measurement failed to stabilize.
Sensors fouled
Clean sensors
during the calibration
Sensors too slow (aged sensor)
Replace sensors
E2
Asymmetry potential too high.
Sensors are aged or polluted
Check buffer solution
(Limits set in service code 21.)
Mistake in calibration
Recalibrate at pH7
Replace
sensor
E3
Slope (sensitivity) is outside limits.
Measuring sensor aged
Replace measuring sensor
(Limits set in service code 22.)
Poor insulation at the connector
Replace or dry cables
E4.1 Impedance of input 1 too low.
Measuring sensor broken
Replace measuring sensor
(Limits set in service code 03.)
Damaged or damp connections
Replace or dry cable
E4.2 Impedance of input 2 too low.
Reference sensor broken
Replace reference sensor
(Limits set in service code 04.)
Damaged connections
Replace cables
E5.1 Impedance of input 1 too high.
Measuring sensor disconnected
Check connections
(Limits set in service code 03.)
Sensors not immersed in process
Check process
Liquid earth disconnected
Check connections
E5.2 Impedance of input 2 too high.
Reference sensor fouled
Clean or replace sensor
(Limits set in service code 04.)
Liquid earth disconnected
Check sensor immersion
Insufficient electrolyte
Check electrolyte reservoir
E7
Temperature sensor open
Process too hot or too cold
Check process
> 140°C (or <-10°C for 8k55)
Wrong temperature sensor setting
Check sensor & setting
Temperature sensor damaged
Check connections
E8
Temperature sensor shortened
Process too cold or too hot
Check process
< -30 °C (or > 120 °C for 8k55)
Wrong temperature sensor used
Check sensor & setting
Temperature sensor damaged
Check connections
E9
Measurement out of range (-2 to 16 pH)
Sensors disconnected
Check cabling
Sensor wrongly connected
Check cabling
Sensor(s) defective
Replace sensor(s)
E10 EEPROM write failure
Fault in electronics
Try again, if unsuccessful
contact
Yokogawa
E11 Wash recovery check error
Measuring sensor aged
Replace measuring sensor
(if communication is set to pH201*B in code
Sensor still coated after washing
Check cleaning system
60)
Defective wash system
If needed adjust timings
E12 ORP / rH outside of preset limits
Sensors disconnected or wrongly
Check cabling
connected
E14 No valid calibration data.
Data lost after switching from
Recalibrate
pH to ORP
E15 Cable resistance to temperature
Cable resistance too high
Use Pt1000
Ω
sensor exceeds limit value.
Corroded contacts
Clean and reterminate
Wrong sensor programmed
Reprogram
E16 Call for maintenance interval time exceeded. System not maintained in
Perform maintenance
preset time period
Reset interval
E17 Output span too small < 1pH
Incorrect configuration by user
Reprogram
E18 Table values make no sense
E19 Programmed values outside acceptable limits Incorrect configuration by user
Reprogram
E20 All programmed data lost
Fault in electronics
Contact Yokogawa
Very severe interference
E21 Checksum error
Software problem
Contact Yokogawa
E23 Zero point outside limits
Sensors are aged or polluted
Check buffer solution
Mistake in calibration
Recalibrate at pH7
Replace
sensor
mA
mA
mA
Summary of Contents for PH202G (S)
Page 5: ...IM 12B07D02 01E ...
Page 9: ...IM 12B07D02 01E LT LV PL EST SLO H BG RO M CZ SK ...
Page 13: ...IM 12B07D02 01E 1 4 Introduction ...
Page 86: ...IM 12B07D02 01E Spare parts 9 1 9 SPARE PARTS See Customer Maintenance Parts List ...
Page 96: ...IM 12B07D02 01E 10 10 Appendix ...
Page 116: ...IM 12B07D02 01E 12 4 Appendix ...
Page 120: ...IM 12B07D02 01E 12 8 Appendix ...
Page 124: ...IM 12B07D02 01E 12 12 Appendix ...
Page 128: ...IM 12B07D02 01E 12 16 Appendix ...