IM12B07D02-01E
Parameter setting 5-19
5-3-5. User interface
Code 50
*RET.
When Auto return is enabled, the transmitter reverts to the measuring mode
from anywhere in the configuration menus, when no button is pressed during
the set time interval of 10 minutes.
Code 51
*MODE
The manual impedance check (on demand) can be setup for operation in the
maintenance mode. (Through the closed front cover).
Code 52
*PASS
Passcodes can be set on any or all of the access levels, to restrict access to
the instrument configuration.
Code 53
*Err.4.1
Error message configuration. Two different types of failure mode can be set.
Hard fail gives a steady FAIL flag in the display, A Fail signal is transmitted on
the outputs when enabled in code 32.
Soft fail gives a flashing FAIL flag in the display. The call for maintenance is
a good example of where a SOFT fail is useful. A warning that the regular
maintenance is due, should not be used to shut down the whole measurement.
Code 54
Not used
Code 55
*CALL.M
Call for maintenance is a trigger to signal that the system has been in service
for longer than the set time without calibration. The user can set up to 250 days
as a routine service interval.
Code 56
*DISP
The display resolution can be set to either 0.01pH or 0.1pH. Not applicable to
the ORP (mV) display.
Summary of Contents for PH202G (S)
Page 5: ...IM 12B07D02 01E ...
Page 9: ...IM 12B07D02 01E LT LV PL EST SLO H BG RO M CZ SK ...
Page 13: ...IM 12B07D02 01E 1 4 Introduction ...
Page 86: ...IM 12B07D02 01E Spare parts 9 1 9 SPARE PARTS See Customer Maintenance Parts List ...
Page 96: ...IM 12B07D02 01E 10 10 Appendix ...
Page 116: ...IM 12B07D02 01E 12 4 Appendix ...
Page 120: ...IM 12B07D02 01E 12 8 Appendix ...
Page 124: ...IM 12B07D02 01E 12 12 Appendix ...
Page 128: ...IM 12B07D02 01E 12 16 Appendix ...