![YOKOGAWA PH202G (S) User Manual Download Page 65](http://html1.mh-extra.com/html/yokogawa/ph202g-s/ph202g-s_user-manual_911726065.webp)
IM12B07D02-01E
5-14 Parameter setting
Code Display
Function
Function
detail
X
Y
Z
Default
values
Temperature measuring and compensation functions
10 *T.SENS Temperature
sensor
Pt1000
0 0 Pt1000
Pt100
1
3kBalco
2
5k1
3
8k55
4
350
5
6k8
6
PTC10k
7
11
*T.UNIT
Display in °C or °F
°C
0
0
°C
°F
1
12
*T.ADJ
Calibrate temperature Adjust to allow for cable resistance
None
13
*T.COMP
Set temp comp
Compensation for process changes off
0
0
Off
Compensation for process changes on
1
*T.COEF
Adjust process TC
Set for TC in pH per 10 °C
-0.00 pH per
10 °C
14-19
Not
used
Summary of Contents for PH202G (S)