IM12B07D02-01E
Parameter setting 5-1
5. PARAMETER SETTING
5-1. Maintenance mode
Standard operation of the EXA instrument involves use of the maintenance (or operating) mode to set up
some of the parameters.
Access to the maintenance mode is available via the six keys that can be pressed through the flexible
window in the instrument cover. Press the MODE-key once to enter this dialog mode.
Note:
At this stage the user will be prompted for pass code where this has been previously set up in
service code 52 in chapter 5.
Automatic calibration
See “calibration” section 6.
(AUT.CAL)
Manual calibration
See “calibration” section 6.
(MAN.CAL)
Sample calibration
See “calibration” section 6.
(SAMPLE)
Display setting
See “operation” section 4.
(DISP)
Manual impedance check
See “parameter setting” §5-1-4 and §5-3-5 code 51.
(MAN.IMP)
Temperature
Set automatic or manual compensation and adjust manual reading
(when pH is set in section 5 service code 01). See adjustment proce-
dure in §5-1-1.
Set automatic reading (when ORP is set in Section 5, service code 01).
See adjustment procedure §5-1-2.
Hold
Manually switch on/off HOLD (when enabled in commissioning menu
section). See adjustment procedure in §5-1-3.
mA
mA
(HOLD)
(TEMP)
Summary of Contents for PH202G (S)
Page 5: ...IM 12B07D02 01E ...
Page 9: ...IM 12B07D02 01E LT LV PL EST SLO H BG RO M CZ SK ...
Page 13: ...IM 12B07D02 01E 1 4 Introduction ...
Page 86: ...IM 12B07D02 01E Spare parts 9 1 9 SPARE PARTS See Customer Maintenance Parts List ...
Page 96: ...IM 12B07D02 01E 10 10 Appendix ...
Page 116: ...IM 12B07D02 01E 12 4 Appendix ...
Page 120: ...IM 12B07D02 01E 12 8 Appendix ...
Page 124: ...IM 12B07D02 01E 12 12 Appendix ...
Page 128: ...IM 12B07D02 01E 12 16 Appendix ...