IM 12B07D02-01E
2-14 Specification
2-11. Control Drawing of PH202S FF/PB Specification (FM Intrinsically safe Entity).
Unclassified Location
Classified Location
Division 1
FM Approved
barrier
Voc (Vt)
d
24 V
Ioc (It)
d
250 mA
Poc (Pt)
d
1.2 W
Ca
t
220pF+ Ccable
La
t
0 H + Lcable
Sensor
Connections
Max. cablelength: 60 mtr.
Cable dia. : 3…12 mm.
FM Class I, DIV. 1, Group ABCD
T4 for ambient temp.
d
55
q
C
-
+
I.S.
certified
Terminator
I.S.
certified
Terminator
PH202S-F
or PH202S-P
Transmitter
Transmitter
-
+
Sensor
Connections
-
+
-
+
x
Sensor(s) are of a passive type to be regarded as 'simple apparatus', devices which neither
store nor generate voltages over 1.5 V, currents over 0.1 A, power over 25 mW or energy over
20
P
J, or are FM Approvals entity approved and meet connection requirements.
x
Electrical data of the PH202S-F & PH202S-P:
-
Supply
circuit:
Maximum input voltage Vmax = 24 V
Maximum input current Imax = 250 mA
Maximum input power Pi=1.2 W
Effective internal capacitance Ci = 220 pF;
Effective internal inductance Li = 0
P
H.
-
Sensor input circuit:
Maximum output voltage Vt = 14.4 V;
Maximum output current It = 32.3 mA
Maximum allowed external capacitance Ca = 600 nF
Maximum allowed external inductance La = 34 mH
x
Any FM Approved barrier may be used that meets the following requirements:
Voc or Vt
d
24 V
Ioc or It
d
250 mA
Poc or Pt
d
1.2 W
Ca
t
220 pF + Ccable; La
t
0
μ
H + Lcable
When installing this equipment, follow the manufacturer’s installation drawing.
Installation should be in accordance with ANSI/ISA RP 12.06.01 “Installation of Intrinsically
Safe Systems for Hazardous (Classified) Locations” and the National Electrical Code
(ANSI/NFPA 70).
Associated apparatus connected to the barrier must not use or generate more than
250 Vrms or Vdc.
x
Resistance between Intrinsically Safe Ground and earth ground must be less than 1.0 Ohm.
x
In case of using cable glands in Outdoor location, they shall be UV rated or made of metal.
Summary of Contents for PH202G (S)
Page 5: ...IM 12B07D02 01E ...
Page 9: ...IM 12B07D02 01E LT LV PL EST SLO H BG RO M CZ SK ...
Page 13: ...IM 12B07D02 01E 1 4 Introduction ...
Page 86: ...IM 12B07D02 01E Spare parts 9 1 9 SPARE PARTS See Customer Maintenance Parts List ...
Page 96: ...IM 12B07D02 01E 10 10 Appendix ...
Page 116: ...IM 12B07D02 01E 12 4 Appendix ...
Page 120: ...IM 12B07D02 01E 12 8 Appendix ...
Page 124: ...IM 12B07D02 01E 12 12 Appendix ...
Page 128: ...IM 12B07D02 01E 12 16 Appendix ...