IM 12B07D02-01E
2-4 Specification
mA-HART® communication
A. Input
: Two wire system 4-20 mA
B. Power supply
:
PH202G :
up to 40 volts
PH202S :
up to 31.5 volts
Note:
The transmitter contains a switched
power supply, drawing its energy
from the 0-4 mA section of the signal.
Consequently the 17 volt limit is
applied at 4 mA. The characteristic of
the unit is such that above about 7 mA
on the output, the terminal voltage can
drop to 14.5 volts without problem.
(see figure 2-2)
C. Transmission:
Isolated output of 4 to 20 mA DC.
D. Signal
: Maximum load 425
Ω
. (see
fi
gure 2-1)
Burn to signal failure acc.
NAMUR Recommendation NE43
(18.01.1994)
E. Operating range :
3.9 to 21mA
F. Communication
:
HART®, 1200 Baud, FSK
modulated on 4 to 20 mA signal
G. Configuration
: Local with 6 keys
H. Software :
Firmware based on Yokogawa stack.
I. Hardware :
Yokogawa HART® Modem F9197UB
J. Other Control systems
:
Yokogawa PRM, Rosemount AMS,
Siemens PDM
K. Hand Terminal :
Rosemount HHT 275/375
L. Other control systems:
Yokogawa PRM,
Rosemount AMS, Siemens PDM
M. Output span
:
- pH
: min 1 pH, max 20 pH.
(max 90% zero suppression)
: The instrument is user
programmable for linear or non-
linear pH ranges.
N. Cable specification
:
0.5 mm diameter or 24 AWG over
maximum length of 1500 m
O. DD specification
:
The PH202 Device Description is
available enabling communications
with the Handheld Communicator
and compatible devices.
mA
NEPSI Certification (PH202S-K)
NEPSI Intrinsically Safe Type
Cert No. GYJ081156X
• Applicable Standard:
GB3836.1-2000, GB3836.4-2000
• Type of Protection and Marking Code:
Ex ia IIC T4/T6
• Ambient Temperature :
T6; –10 to 40°C, T4; –10 to 55°C
Note 1 Entity Parameters
• Intrinsically safe input parameters
(te and -):
Maximum Input Voltage (Ui) = 31.5 V
Maximum Input Current (Ii) = 100 mA
Maximum Input Power (Pi) = 1.2 W
Maximum Internal Capacitance (Ci) = 22 nF
Maximum Internal Inductance (Li) = 35
μ
H
• Intrinsically safe output parameters and
maximum external parameters
(terminal 11 and 17):
Uo=14.4 V, Io=32.3 mA, Po=0.12 W, Co=600
nF, Lo=34 mH
Note 2 Installation
• Electrostatic charges on the display window shall
be avoided.
• The external earth connection facility shall be
connected reliably.
• The instrument modification or parts replacement
by other than authorized representative of
Yokogawa Electric Corporation and will void
NEPSI Intrinsically safe certification.
• The user shall not change the configuration in
order to maintain/ensure the explosion protection
performance of the equipment. Any change may
impair safety.
• For installation, use and maintenance of
the product, the end user shall observe the
instruction manual and the following standards:
GB50257-1996 "Code for construction and
acceptance of electric device for explosion
atmospheres and fire hazard electrical
equipment installation engineering''.
GB3836.13-1997 "Electrical apparatus for
explosive gas atmospheres Part 13: Repair and
overhaul for apparatus used in explosive gas
atmospheres".
GB3836.15-2000 "Electrical apparatus for
explosive gas atmospheres- Part 15: Electrical
installations in hazardous area (other than
mines)" .
GB3836.16-2006 "Electrical apparatus for
explosive gas atmospheres- Part 16: lnspection
and maintenance of electrical installation (other
than mines)".
mA
Summary of Contents for PH202G (S)
Page 5: ...IM 12B07D02 01E ...
Page 9: ...IM 12B07D02 01E LT LV PL EST SLO H BG RO M CZ SK ...
Page 13: ...IM 12B07D02 01E 1 4 Introduction ...
Page 86: ...IM 12B07D02 01E Spare parts 9 1 9 SPARE PARTS See Customer Maintenance Parts List ...
Page 96: ...IM 12B07D02 01E 10 10 Appendix ...
Page 116: ...IM 12B07D02 01E 12 4 Appendix ...
Page 120: ...IM 12B07D02 01E 12 8 Appendix ...
Page 124: ...IM 12B07D02 01E 12 12 Appendix ...
Page 128: ...IM 12B07D02 01E 12 16 Appendix ...