IM 12B07D02-01E
Specification 2-11
2-8. Control Drawing of PH202S mA HART® Specification (CSA).
S a fe a re a
V m ax = 3 1 .5 V o ltD C
Im ax = 1 0 0 m A
P m ax = 1 .2 W att
C S A certified
P ow e r S upp ly
(H A R T c o m pa tib le ) )
O utput
S upp ly
S en sor(s)
term in als 1 1 -1 7
+
_
G
S en sor(s)
term in a ls 1 1 -1 7
H azard ous area
S a fe are a
+
_
G
Func tional
e a rth
e a rth
Intrins ic a lly sa fe de sign
(p H /O R P -transm itter)
P H 20 2S
+
_
Load
R esistance
Im ax = 1 0 0 m A
V m ax = 3 1 .5 V o ltD C
H azard ous area
e arth
+
_
24 volts D C N o m ina l
S upp ly V oltage.
Intrins ic a lly sa fe de sign
C S A E x ia C la ss I, D iv.1 , G roup A B C D
T 4 for a m bient te m p . < 5 5°C
T 6 fo r am b ie nt te m p . < 4 0°C
(p H /O R P -transm itter)
P H 2 02 S
C S A certified
safety barrier o r po w er sup ply
S u itable valu es are:
(H A R T com patible )
C S A E x ia C lass I, D iv.1 , G roup A B C D
T 4 fo r am b ient tem p . < 5 5°C
T 6 fo r am b ient tem p . < 4 0°C
Fo r electrical d ata:
see text belo w .
Fo r electrical d ata:
see text belo w .
S uitab le values are:
Functio na l
Func tional
・
Sensor(s) are thermocouples, RTD’s, passive resistive switch devices, or are CSA entity approved
and meet connection requirements.
・
Electrical data of the PH202S.
- Supply and output circuit (ter and -):
Maximum input voltage Vmax = 31.5 V. Maximum input current Imax = 100 mA.
Maximum input power Pmax = 1.2 W.
Effective internal capacitance Ci = 22 nF. Effective internal inductance Li = 35
P
H.
- Sensor input circuit (terminals 11 through 17):
Maximum output voltage Voc = 14.4 V. Maximum output current Isc = 32.3 mA.
Maximum allowed external capacitance Ca = 600 nF
Maximum allowed external inductance La = 34 mH.
・
Barriers and power supply should be CSA certified. The specifications must not exceed
the maximum values as shown in the diagram above.
Installation should be in accordance with Canadian Electrical Code, Part I.
Maximum safe area voltage should not exceed 250 VRMS.
For Class I, Div. 2, Group ABCD the CSA certified barrier is not required, and the Sensor
input circuit (terminals 11 through 17) is non-incendive having the parameters :
Maximum output voltage Voc = 14.4 V. Maximum output current Isc = 32.3 mA.
Maximum allowed external capacitance Ca = 3.5
P
F.
Maximum allowed external inductance La = 76 mH.
・
The Hand Held Communicator must be of a CSA certified intrinsically safe type
in case it is used on the intrinsically safe circuit in the hazardous area, or of a
CSA certified
non-incendive type in case it is used on the non-incendive circuit in the
hazardous area.
Summary of Contents for PH202G (S)
Page 5: ...IM 12B07D02 01E ...
Page 9: ...IM 12B07D02 01E LT LV PL EST SLO H BG RO M CZ SK ...
Page 13: ...IM 12B07D02 01E 1 4 Introduction ...
Page 86: ...IM 12B07D02 01E Spare parts 9 1 9 SPARE PARTS See Customer Maintenance Parts List ...
Page 96: ...IM 12B07D02 01E 10 10 Appendix ...
Page 116: ...IM 12B07D02 01E 12 4 Appendix ...
Page 120: ...IM 12B07D02 01E 12 8 Appendix ...
Page 124: ...IM 12B07D02 01E 12 12 Appendix ...
Page 128: ...IM 12B07D02 01E 12 16 Appendix ...