Automatic Calibration With Hold Active
Note: To start calibration with another solution, press NO to cycle through the other options.
"CAL 7"
NO
"CAL 4"
NO
"CAL 9"
NO
"CAL 7"
The single-
point
calibration
is now
complete.
put sensors
back in the
process and
press
NO
to
switch off
HOLD and
return to
measuring
mode.
Transfer to pH 4
buffer and press
YES
.
The instrument then waits for the reading
to stabilise. (The pH display flashes).
When the reading is stable, the CAL END
message appears. Press
YES
.
WAIT flashes briefly then HOLD
The calibration is now complete. Put the sensors back in the process
and press
NO
to turn off HOLD and return to the measuring mode.
NO
Press
YES
for single point
(As Pot) adjustment.
Press
NO
to proceed to calibrate Slope.
The instruments waits for
the reading to stabilise.
(The pH display flashes)
When reading Is stable,
the CAL END message
appears.
YES
Press the
MODE
key. AUT.CAL appears in the display, and the
YES/NO
key prompt flags flash. Press
YES
.
NEW.SNS: Display flashes
YES/NO
answer
YES
if new sensor fitted or
NO
if not.
Care! YES resets logbook calibration data.
Insert the sensors in pH 7 buffer solution.
Press
YES
to start calibration.
pH
ENT
MAN.CAL
DISPLAY
HOLD
NO
MODE
YES
YOKOGAWA
MODE
TEMP
AUT.CAL
MEASURE
6-5-2. Automatic calibration with HOLD active
mA
Summary of Contents for PH202G (S)
Page 5: ...IM 12B07D02 01E ...
Page 9: ...IM 12B07D02 01E LT LV PL EST SLO H BG RO M CZ SK ...
Page 13: ...IM 12B07D02 01E 1 4 Introduction ...
Page 86: ...IM 12B07D02 01E Spare parts 9 1 9 SPARE PARTS See Customer Maintenance Parts List ...
Page 96: ...IM 12B07D02 01E 10 10 Appendix ...
Page 116: ...IM 12B07D02 01E 12 4 Appendix ...
Page 120: ...IM 12B07D02 01E 12 8 Appendix ...
Page 124: ...IM 12B07D02 01E 12 12 Appendix ...
Page 128: ...IM 12B07D02 01E 12 16 Appendix ...