IM 12B07D02-01E
Installation and wiring 3-3
3-2. Preparation
Refer to figure 3-4. The power/output connections and the sensor connections should be made in
accordance with the diagram on page 3-6. The terminals are of a plug in style for ease of mounting.
To open the EXA 202 for wiring:
1. Loosen the four frontplate screws and remove the cover.
2. The terminal strip is now visible.
3.
Connect the power supply. Use the gland on the left for this cable.
4.
Connect the sensor input, using the gland on the right (see fig. 3-5). Switch on the power.
Commission the instrument as required or use the default settings.
5. Replace the cover and secure frontplate with the four screws.
6. Connect the grounding terminals tp protective earth.
7.
The optional hose connection is used to guide the cables coming from an immersion fitting through a
protective plastic tubing to the transmitter.
3-2-1. Cables, terminals and glands
The PH202 is equipped with terminals suitable for the connection of finished cables in the size range:
0.13 to 2.5 mm (26 to 14 AWG). The glands will form a tight seal on cables with an outside diameter in
the range of 6 to 12 mm (0.24 to 0.47 inches).
Figure 3-5. Glands to be used for cabling
Grounding
terminal
Power/Output
cable gland
Sensor
cable gland
Summary of Contents for PH202G (S)
Page 5: ...IM 12B07D02 01E ...
Page 9: ...IM 12B07D02 01E LT LV PL EST SLO H BG RO M CZ SK ...
Page 13: ...IM 12B07D02 01E 1 4 Introduction ...
Page 86: ...IM 12B07D02 01E Spare parts 9 1 9 SPARE PARTS See Customer Maintenance Parts List ...
Page 96: ...IM 12B07D02 01E 10 10 Appendix ...
Page 116: ...IM 12B07D02 01E 12 4 Appendix ...
Page 120: ...IM 12B07D02 01E 12 8 Appendix ...
Page 124: ...IM 12B07D02 01E 12 12 Appendix ...
Page 128: ...IM 12B07D02 01E 12 16 Appendix ...