IM 12B07D02-01E
Installation and wiring 3-7
3-5. Wiring the sensor system
3-5-1. Impedance measurement jumper settings
NOTE:
It is important to decide first which application and which settings are appropriate for the
installation. This decision is best made before the jumpers are installed, because the cables will
rest beside the jumpers in their installed positions.
Table 3-1. Impedance measuring jumpers
Figure no.
Jumper Settings
Jumper Settings
Application & Sensor Connections
Input #1
Input #2
1
High Impedance
Low Impedance
Normal pH sensors
Glass
sensor
on
Input
#1
Reference
sensor
on
Input
#2
2
High Impedance
High Impedance
Special electrodes using
2
glass
sensors
(e.g.
Pfaudler
18)
3
Low Impedance
High Impedance
ORP (pH compensated) and/or rH
metal
sensor
on
Input
#1
pH
glass
(as
reference)
on
Input
#2
4
Low Impedance
Low Impedance
ORP (Redox measurement)
metal
sensor
on
Input
#1
Normal
reference
on
Input
#2
For convenience insulated jumper links are provided. Ordinary wire links can also be used, and are just
as effective.
The following four jumper figure illustrations (figure 3-8) show the jumper positions related to the figure
numbers in the above table.
11 12 14 17 13
15 16
HIGH IMP
LOW IMP
SUPPLY
TEMP LE
INPUT
2
INPUT
1
11 12 14 17 13
15 16
HIGH IMP
LOW IMP
SUPPLY
TEMP LE
INPUT
2
INPUT
1
11 12 14 17 13
15 16
HIGH IMP
LOW IMP
SUPPLY
TEMP LE
INPUT
2
INPUT
1
11 12 14 17 13
15 16
HIGH IMP
LOW IMP
SUPPLY
TEMP LE
INPUT
2
INPUT
1
1
2
3
4
Setting on shipment
Fig. 3-8. Jumper positions
Summary of Contents for PH202G (S)
Page 5: ...IM 12B07D02 01E ...
Page 9: ...IM 12B07D02 01E LT LV PL EST SLO H BG RO M CZ SK ...
Page 13: ...IM 12B07D02 01E 1 4 Introduction ...
Page 86: ...IM 12B07D02 01E Spare parts 9 1 9 SPARE PARTS See Customer Maintenance Parts List ...
Page 96: ...IM 12B07D02 01E 10 10 Appendix ...
Page 116: ...IM 12B07D02 01E 12 4 Appendix ...
Page 120: ...IM 12B07D02 01E 12 8 Appendix ...
Page 124: ...IM 12B07D02 01E 12 12 Appendix ...
Page 128: ...IM 12B07D02 01E 12 16 Appendix ...