IM 12B07D02-01E
3-4 Installation and wiring
Figure 3-6. System configuration
3-3. Wiring of sensors
3-3-1. General precautions
Generally, transmission of signals from pH sensors is at a very low voltage and high impedance level.
Thus a lot of care must be taken to avoid interference. Before connecting sensor cables to the transmit-
ter make sure that next conditions are met:
– the sensor cables are not mounted in tracks together with high voltage and or power switching
cables
– only standard coaxial electrode cables or extension cable are used
– the transmitter is mounted within the distance of the sensor cables (max. 10 m)
– the setup is kept flexible for easy insertion and retraction of the sensors in the fitting.
3-3-2. Additional precautions for installations in hazardous areas
Make sure that the total of capacitance and inductances connected to the input terminals of the EXA
PH202S do not exceed the limits given in the certificate.
This sets a limit to the cable and extensions used.
– The intrinsic safe version of the PH202S instrument can be mounted in Zone 0 or 1 (PH202S-B, -N,
-D).
– The sensors can be installed in Zone 0 or Zone 1 if a safety barrier according to the limits given in
the system certificate is used.
– Ensure that the total of capacitance and inductances connected to the terminals of the EXA
PH202S do not exceed the limits given in the certificate of the safety barrier or distributor.
– The cable used should preferably have a BLUE colour or marking on the outside.
– Installation for (sensors in Zone 0 or 1):
Generally, the distributor with input/output isolation has no external earth connection. If there is an
earth connection on the distributor and the external connection of the transmitter is connected to
“protective” earth, the shield of the 2-wire cable may NOT be connected to “protective” earth at the
distributor too.
1 2
1 8 0
1 0 0
0
NO
MODE
YES
>
ENT
>
F AIL
HOLD
YES NO
ENT
MEASURE
MAN.CAL
DISPLAY
HOLD
YOKOGAWA
MODE
TEMP
AUT.CAL
TEMP.MAN.
pH
Sensors
Recorder
Computer
Hand Held
Communicator
Output/supply
Input
2 , 5 or 10 m
Safety Barrier
PH202S only
CURRENT OUTPUT
ref pH/ORP emp.
Distributor
Summary of Contents for PH202G (S)
Page 5: ...IM 12B07D02 01E ...
Page 9: ...IM 12B07D02 01E LT LV PL EST SLO H BG RO M CZ SK ...
Page 13: ...IM 12B07D02 01E 1 4 Introduction ...
Page 86: ...IM 12B07D02 01E Spare parts 9 1 9 SPARE PARTS See Customer Maintenance Parts List ...
Page 96: ...IM 12B07D02 01E 10 10 Appendix ...
Page 116: ...IM 12B07D02 01E 12 4 Appendix ...
Page 120: ...IM 12B07D02 01E 12 8 Appendix ...
Page 124: ...IM 12B07D02 01E 12 12 Appendix ...
Page 128: ...IM 12B07D02 01E 12 16 Appendix ...