5-18 Parameter setting
Code Display
Function
Function detail
X
Y
Z
Default values
mA Outputs
30
Not
used
31
*OUTP.F
mA output functions
pH
0
0
Code 01 set for pH
pH (table)
1
Parameter 2 (with suitable sensor(s),
2
and when enabled in code 02)
32
*BURN
Burn function
No burnout
0
0 No Burn.
Burnout downscale
1
Burnout upscale
2
Pulse burnout
3
33, 34
Not used
35
*TABLE
Output table for mA
*0%
Linearisation table for mA1 in 5% steps.
*5%
The measured value is set in the main
*10%
display using the >, ^, ENT keys, for
...
each of the 5% interval steps.
...
Where a value is not known, that value may
*90%
be skipped, and a linear interpolation will
*100%
take
place.
36-39
Not
used
mA