IM 12B07D02-01E
7-2 maintenance
7-3. Calibration procedures are described in step-by-step detail in chapter 6. However, follow these
guidelines.
1.
Before starting a calibration, make sure the electrode system is properly cleaned so that electrodes
are fully functional. They must then be rinsed with clean water to avoid contamination of the
calibration solution.
2.
Always use fresh buffer solutions to avoid the possibility of introducing errors from contaminated or
aged solutions. Buffers supplied as liquids have a limited shelf life, especially alkaline buffers which
absorb CO
2
from the air.
3.
Yokogawa strongly recommends NIST (primary) buffer standards in order to ensure the best
accuracy and best buffer capacity is available. Commercially adjusted buffers (e.g. 7.00, 9.00
or 10.00 pH) are a compromise as a standard, and are often supplied without the temperature
dependency curve. Their stability will be much worse than for NIST solutions.
NOTE
:
NIST (formerly NBS) buffers are available as consumable items from any Yokogawa sales office
under the following part numbers:
6C232 4.01 pH at 25°C}
6C237 6.87 pH at 25°C} A box contains 5 packets of powder. Each makes a 200 ml solution.
6C236 9.18 pH at 25°C}
Summary of Contents for PH202G (S)
Page 5: ...IM 12B07D02 01E ...
Page 9: ...IM 12B07D02 01E LT LV PL EST SLO H BG RO M CZ SK ...
Page 13: ...IM 12B07D02 01E 1 4 Introduction ...
Page 86: ...IM 12B07D02 01E Spare parts 9 1 9 SPARE PARTS See Customer Maintenance Parts List ...
Page 96: ...IM 12B07D02 01E 10 10 Appendix ...
Page 116: ...IM 12B07D02 01E 12 4 Appendix ...
Page 120: ...IM 12B07D02 01E 12 8 Appendix ...
Page 124: ...IM 12B07D02 01E 12 12 Appendix ...
Page 128: ...IM 12B07D02 01E 12 16 Appendix ...